Operator's Manual Part 1

Table Of Contents
Revision D ApexPro Telemetry System 4-7
2001989-134
Connection: Apex Oximeter Setup
Apex Oximeter Setup
Apex Oximeter Battery Installation
Install two new AA alkaline batteries in the Apex Oximeter:
1. Locate the battery cover at the bottom of the back of the Apex
Oximeter.
2. Press the latch tab and lift up to open the battery compartment.
3. Insert the batteries, being careful to follow the polarity signs located
within the battery compartment.
4. Close the battery cover.
NOTE
When the digital displays on the Apex Oximeter start flashing, the
Apex Oximeter has approximately one hour of reserve power left
before the unit shuts down.
Battery Functional Life
The Apex Oximeter runs on two AA batteries. Battery life is
approximately 60 hours.
For optimum performance, follow these guidelines:
Install two new alkaline batteries each time you begin monitoring a
new patient.
Install two new alkaline batteries whenever the digital displays on
the Apex Oximeter start flashing.
Always change both batteries at the same time.
Always use new batteries.
CAUTION
GE Medical Systems Information Technologies
recommends that you always replace both batteries at
the same time. Re-using old batteries or using a
combination of old and new batteries in the Apex
Oximeter will compromise functionality of the
transmitter and increase the risk of fire hazard.