Operator's Manual Part 1

Table Of Contents
3-10 ApexPro Telemetry System Revision D
2001989-134
Equipment Overview: ApexPro Telemetry System
ApexPro CH Transmitter (not for sale outside of the U.S. and Canada)
The Channel Number (CH) is how the ApexPro CH transmitter is
identified. During setup, when the TTX ID is entered for a patient at the
CIC Pro, the CIC Pro recognizes the transmitter type and translates the
information into an alpha-numeric number.
The alpha-numeric number of the transmitter is displayed under the
ECG parameter window. It identifies the type of transmitter (AP for
ApexPro or CH for ApexPro CH transmitters), and the channel number
for ApexPro CH transmitters
The ApexPro CH transmitter is designed to be IPX7 compliant, so it can
survive inadvertent submersion. It supports the same features as the
ApexPro transmitter with the exception of the Event Marker button.
AperPro CH Transmitter
Event Marker
Pressing the Event Marker button, located on the front panel of the
ApexPro CH transmitter, marks a patient event.
The CIC Pro responds to an event marker by displaying a blue border
around the event bed and sounding an alarm tone. The message Remote
Event appears under the ECG parameter window for approximately ten
seconds.
When enabled, the event marker will generate a 20-second graph and an
event in alarm history. The graph feature can be turned off in the Setup
CIC tab sheet, Event Marker Graph On/Off. The event marker can be
disabled on the Setup CIC tab sheet, Event Marker On/Off.
Verify Leads
Checks the lead/
skin preparation
quality.
Graph – Initiates
the printing of a
graph strip.
Pause Alarm LED
Good lead LEDs
Change Battery
LED
Dust covers and
INTFC connector
ports
432A
Event Marker
Marks a patient
event.