Admit/View a Patient: About Admitting
Combo Monitoring
This application provides the option to acquire ECG data from either the
monitor or from a telemetry receiver system. This ECG data acquisition
capability enhances basic telemetry monitoring by providing additional
access to all of the available parameters from the monitor. A Unity
network connection is required.
NOTE
Users should be aware of a possible time discrepancy between the
waveforms from the Telemetry device and the waveforms hard-wired
to the display device. Users should not consider these waveforms to
be synchronous. If absolute synchronicity is desired, Combo mode
should be discontinued and the ECG waveforms should be acquired
via the hard-wired bedside device.
Locked and Unlocked Beds
The CIC Pro can be configured with the bed names in either locked or
unlocked mode. When locked, the bed names are permanently assigned
to specific windows.
For information about locked and unlocked beds, refer to the service
manual for the CIC Pro.
NOTE
It is possible to admit a patient to a window with a bed name that is
locked to NONE. To avoid duplication of patient waveforms, a
window locked to NONE should not be used to admit a patient.
Alpha-Numeric TTX ID Number
Each transmitter has a transmitter (TTX) ID that corresponds to a
frequency and service type. The TTX ID is found in parenthesis on the
back of the ApexPro transmitter.
When the TTX ID is entered for a patient at the CIC Pro, the CIC Pro
recognizes the transmitter type and translates the information into an
alpha-numeric number. The alpha-numeric number of the transmitter is
displayed under the ECG parameter window.
NOTE
The TTX ID number is composed of either a three, four or five digit
number.