Operator's Manual Part 1
Table Of Contents
- Contents
- 1 The Basics 1-1
- 2 Safety 2-1
- 3 Equipment Overview 3-1
- Introduction 3-3
- ApexPro Telemetry System 3-4
- Compatibility with Bedside Monitors 3-5
- ApexPro Transmitters 3-6
- ApexPro CH Transmitter (not for sale outside of the U.S. and Canada) 3-10
- ApexPro FH Transceiver (not for sale outside of the U.S.) 3-11
- Apex Oximeter SpO2 Module 3-12
- Xpod™ Oximeter 3-13
- Accutracker DX Noninvasive Blood Pressure (NBP) Monitor 3-14
- DINAMAP® PRO Series Monitors 3-16
- Antenna System 3-17
- Receiver System 3-17
- Unity Network 3-17
- CIC Pro Clinical Information Center 3-18
- 4 Connection 4-1
- 5 Maintenance 5-1
- 6 Telemetry Setup 6-1
- 7 Admit/View a Patient 7-1
- 8 Alarm Control 8-1
- 9 Printing 9-1
- 10 Patient Data 10-1
- 11 ECG Monitoring 11-1
- 12 SpO2 Monitoring 12-1
- 13 NBP Monitoring 13-1
- Introduction 13-3
- Safety Considerations 13-4
- Programming the Blood Pressure Monitor 13-5
- Patient Preparation 13-9
- NBP Monitoring 13-13
- Pressures Tab Sheet 13-15
- Troubleshooting 13-17
- Message Glossary A-3
- Contact Information B-3
- Abbreviations and Symbols C-3
- The Basics D-3
- Safety D-5
- Equipment Overview D-8
- Connections D-14
- Maintenance D-18
- Compliance D-26
- 1 The Basics
- 2 Safety
- 3 Equipment Overview
- Introduction
- ApexPro Telemetry System
- Compatibility with Bedside Monitors
- ApexPro Transmitters
- ApexPro CH Transmitter (not for sale outside of the U.S. and Canada)
- ApexPro FH Transceiver (not for sale outside of the U.S.)
- Apex Oximeter SpO2 Module
- Xpod™ Oximeter
- Accutracker DX Noninvasive Blood Pressure (NBP) Monitor
- DINAMAP® PRO Series Monitors
- Antenna System
- Receiver System
- Unity Network
- CIC Pro Clinical Information Center
- 4 Connection
- 5 Maintenance
- 6 Telemetry Setup
- 7 Admit/View a Patient
Revision D ApexPro Telemetry System 6-15
2001989-134
Telemetry Setup: CIC Defaults
The following features of the CIC Defaults tab sheet can be modified in
the service mode.
Name:
Central—allows you to enter the name for this CIC Pro.
Unit—allows you to enter the care unit name for use on the Unity IS
network.
Waveforms:
ECG 1—this is defined by the ECG source.
Waveform 2 – 4—allows you to define multiple patient viewer
waveforms.
Mirror Central Display—allows double-monitoring of patients from
remote, or secondary, central stations.
Browser Configuration—allows you to browse your intranet using the
provided web browser.
Alarm Volume—allows you to set a default alarm volume.
Printer/Writer:
Laser—allows you to designate the default laser printer for this CIC
Pro.
DDW—allows you to designate the default writer for this CIC Pro.
Full Disclosure—allows you to designate the default printer to print
Full Disclosure for this clinical information center.
Cancel Print Jobs—This control allows you to cancel a print job to the
Laser printer and Full Disclosure printer.
Color Set:
Clinical—selects a preset clinical color scheme for waveforms.
Transducer—selects a preset transducer color scheme for waveforms.
Custom—allows you to set each waveform color individually.