Operator's Manual Part 1
Table Of Contents
- Contents
- 1 The Basics 1-1
- 2 Safety 2-1
- 3 Equipment Overview 3-1
- Introduction 3-3
- ApexPro Telemetry System 3-4
- Compatibility with Bedside Monitors 3-5
- ApexPro Transmitters 3-6
- ApexPro CH Transmitter (not for sale outside of the U.S. and Canada) 3-10
- ApexPro FH Transceiver (not for sale outside of the U.S.) 3-11
- Apex Oximeter SpO2 Module 3-12
- Xpod™ Oximeter 3-13
- Accutracker DX Noninvasive Blood Pressure (NBP) Monitor 3-14
- DINAMAP® PRO Series Monitors 3-16
- Antenna System 3-17
- Receiver System 3-17
- Unity Network 3-17
- CIC Pro Clinical Information Center 3-18
- 4 Connection 4-1
- 5 Maintenance 5-1
- 6 Telemetry Setup 6-1
- 7 Admit/View a Patient 7-1
- 8 Alarm Control 8-1
- 9 Printing 9-1
- 10 Patient Data 10-1
- 11 ECG Monitoring 11-1
- 12 SpO2 Monitoring 12-1
- 13 NBP Monitoring 13-1
- Introduction 13-3
- Safety Considerations 13-4
- Programming the Blood Pressure Monitor 13-5
- Patient Preparation 13-9
- NBP Monitoring 13-13
- Pressures Tab Sheet 13-15
- Troubleshooting 13-17
- Message Glossary A-3
- Contact Information B-3
- Abbreviations and Symbols C-3
- The Basics D-3
- Safety D-5
- Equipment Overview D-8
- Connections D-14
- Maintenance D-18
- Compliance D-26
- 1 The Basics
- 2 Safety
- 3 Equipment Overview
- Introduction
- ApexPro Telemetry System
- Compatibility with Bedside Monitors
- ApexPro Transmitters
- ApexPro CH Transmitter (not for sale outside of the U.S. and Canada)
- ApexPro FH Transceiver (not for sale outside of the U.S.)
- Apex Oximeter SpO2 Module
- Xpod™ Oximeter
- Accutracker DX Noninvasive Blood Pressure (NBP) Monitor
- DINAMAP® PRO Series Monitors
- Antenna System
- Receiver System
- Unity Network
- CIC Pro Clinical Information Center
- 4 Connection
- 5 Maintenance
- 6 Telemetry Setup
- 7 Admit/View a Patient
Revision D ApexPro Telemetry System 6-13
2001989-134
Telemetry Setup: Telemetry Unit Defaults
Transmitter Alarm Pause
NOTE
Refer to Chapter 8, Alarm Control, for important information about
Transmitter Alarm Pause.
The options found in the popup list for the Transmitter Alarm Pause field
are:
Disabled (factory default selection)
Enabled
Off
This option sets the unit default for all telemetry patients admitted to
the CIC Pro. When the Transmitter Alarm Pause option is set to
Disabled or Enabled, transmitter alarm pause privileges can also be set
on an individual patient basis using the Enable Transmitter Pause check
box in the patient’s Alarm Control tab sheet.
Alarm Pause Breakthrough
NOTE
Refer to Chapter 8, Alarm Control, for important information about
Alarm Pause Breakthrough.
The Alarm Pause Breakthrough option allows you to set the default for
when a crisis alarm will sound, even if an alarm pause has been set. The
choices are:
Always On (factory default selection)
Always Off
This option sets the unit default for all telemetry patients admitted to
the CIC Pro. It cannot be changed on an individual patient basis.
Event Marker
The Event Marker option allows you to enable/disable this feature, which
is used to remotely mark a patient’s event (refer to “Event Marker” on
page 3-10). The choices are:
Off (factory default selection)
On
This feature is not applicable to all transmitters.