Telemetry Setup: Telemetry Unit Defaults
Transmitter Alarm Pause
NOTE
Refer to Chapter 8, Alarm Control, for important information about
Transmitter Alarm Pause.
The options found in the popup list for the Transmitter Alarm Pause field
are:
Disabled (factory default selection)
Enabled
Off
This option sets the unit default for all telemetry patients admitted to
the CIC Pro. When the Transmitter Alarm Pause option is set to
Disabled or Enabled, transmitter alarm pause privileges can also be set
on an individual patient basis using the Enable Transmitter Pause check
box in the patient’s Alarm Control tab sheet.
Alarm Pause Breakthrough
NOTE
Refer to Chapter 8, Alarm Control, for important information about
Alarm Pause Breakthrough.
The Alarm Pause Breakthrough option allows you to set the default for
when a crisis alarm will sound, even if an alarm pause has been set. The
choices are:
Always On (factory default selection)
Always Off
This option sets the unit default for all telemetry patients admitted to
the CIC Pro. It cannot be changed on an individual patient basis.
Event Marker
The Event Marker option allows you to enable/disable this feature, which
is used to remotely mark a patient’s event (refer to “Event Marker” on
page 3-10). The choices are:
Off (factory default selection)
On
This feature is not applicable to all transmitters.