Operator's Manual Part 1
Table Of Contents
- Contents
- 1 The Basics 1-1
- 2 Safety 2-1
- 3 Equipment Overview 3-1
- Introduction 3-3
- ApexPro Telemetry System 3-4
- Compatibility with Bedside Monitors 3-5
- ApexPro Transmitters 3-6
- ApexPro CH Transmitter (not for sale outside of the U.S. and Canada) 3-10
- ApexPro FH Transceiver (not for sale outside of the U.S.) 3-11
- Apex Oximeter SpO2 Module 3-12
- Xpod™ Oximeter 3-13
- Accutracker DX Noninvasive Blood Pressure (NBP) Monitor 3-14
- DINAMAP® PRO Series Monitors 3-16
- Antenna System 3-17
- Receiver System 3-17
- Unity Network 3-17
- CIC Pro Clinical Information Center 3-18
- 4 Connection 4-1
- 5 Maintenance 5-1
- 6 Telemetry Setup 6-1
- 7 Admit/View a Patient 7-1
- 8 Alarm Control 8-1
- 9 Printing 9-1
- 10 Patient Data 10-1
- 11 ECG Monitoring 11-1
- 12 SpO2 Monitoring 12-1
- 13 NBP Monitoring 13-1
- Introduction 13-3
- Safety Considerations 13-4
- Programming the Blood Pressure Monitor 13-5
- Patient Preparation 13-9
- NBP Monitoring 13-13
- Pressures Tab Sheet 13-15
- Troubleshooting 13-17
- Message Glossary A-3
- Contact Information B-3
- Abbreviations and Symbols C-3
- The Basics D-3
- Safety D-5
- Equipment Overview D-8
- Connections D-14
- Maintenance D-18
- Compliance D-26
- 1 The Basics
- 2 Safety
- 3 Equipment Overview
- Introduction
- ApexPro Telemetry System
- Compatibility with Bedside Monitors
- ApexPro Transmitters
- ApexPro CH Transmitter (not for sale outside of the U.S. and Canada)
- ApexPro FH Transceiver (not for sale outside of the U.S.)
- Apex Oximeter SpO2 Module
- Xpod™ Oximeter
- Accutracker DX Noninvasive Blood Pressure (NBP) Monitor
- DINAMAP® PRO Series Monitors
- Antenna System
- Receiver System
- Unity Network
- CIC Pro Clinical Information Center
- 4 Connection
- 5 Maintenance
- 6 Telemetry Setup
- 7 Admit/View a Patient
6-12 ApexPro Telemetry System Revision D
2001989-134
Telemetry Setup: Telemetry Unit Defaults
Patient Age
The options found in the pull-down list for the Patient Age field are:
0-2 years
3-10 years
11-13 years
Adult (factory default selection)
The patient age setting chosen in the Telemetry Unit Defaults tab sheet
affects the alarm settings. Refer to the example below.
NOTE
This is an example of Brady changes associated with age changes.
The Telemetry Unit Defaults tab sheet settings take precedence
when the patient age chosen in the Admit tab sheet and the age
default setting from the Telemetry Unit Defaults tab sheet match.
When the ages do not match, the Admit tab sheet age setting takes
precedence.
Example 1. Patient Age
Telemetry Unit Defaults
Tab Sheet Setting
Age Chosen In The Admit Tab
Sheet When Patient Admitted Resulting Limits
Unit Default
Alarm Level (Brady)
Resulting
Alarm Level
Adult Adult
0-2 years
3-10 years
11-13 years
10, 155
1
90, 200
60, 180
50, 150
Message
Message
Message
Message
Message
Crisis
Crisis
Advisory
0-2 years Adult
0-2 years
3-10 years
11-13 years
50, 150
10, 155
*
60, 180
50, 150
Message
Message
Message
Message
Advisory
Message
Crisis
Advisory
3-10 years Adult
0-2 years
3-10 years
11-13 years
50, 150
90, 200
10, 155
*
50, 150
Message
Message
Message
Message
Advisory
Crisis
Message
Advisory
11-13 years Adult
0-2 years
3-10 years
11-13 years
50, 150
90, 200
60, 180
10, 155
*
Message
Message
Message
Message
Advisory
Crisis
Crisis
Message
1. These reflect the unit default values set for this example. These values will reflect whatever your unit default is set to.