Operator's Manual Part 1
Table Of Contents
- Contents
- 1 The Basics 1-1
- 2 Safety 2-1
- 3 Equipment Overview 3-1
- Introduction 3-3
- ApexPro Telemetry System 3-4
- Compatibility with Bedside Monitors 3-5
- ApexPro Transmitters 3-6
- ApexPro CH Transmitter (not for sale outside of the U.S. and Canada) 3-10
- ApexPro FH Transceiver (not for sale outside of the U.S.) 3-11
- Apex Oximeter SpO2 Module 3-12
- Xpod™ Oximeter 3-13
- Accutracker DX Noninvasive Blood Pressure (NBP) Monitor 3-14
- DINAMAP® PRO Series Monitors 3-16
- Antenna System 3-17
- Receiver System 3-17
- Unity Network 3-17
- CIC Pro Clinical Information Center 3-18
- 4 Connection 4-1
- 5 Maintenance 5-1
- 6 Telemetry Setup 6-1
- 7 Admit/View a Patient 7-1
- 8 Alarm Control 8-1
- 9 Printing 9-1
- 10 Patient Data 10-1
- 11 ECG Monitoring 11-1
- 12 SpO2 Monitoring 12-1
- 13 NBP Monitoring 13-1
- Introduction 13-3
- Safety Considerations 13-4
- Programming the Blood Pressure Monitor 13-5
- Patient Preparation 13-9
- NBP Monitoring 13-13
- Pressures Tab Sheet 13-15
- Troubleshooting 13-17
- Message Glossary A-3
- Contact Information B-3
- Abbreviations and Symbols C-3
- The Basics D-3
- Safety D-5
- Equipment Overview D-8
- Connections D-14
- Maintenance D-18
- Compliance D-26
- 1 The Basics
- 2 Safety
- 3 Equipment Overview
- Introduction
- ApexPro Telemetry System
- Compatibility with Bedside Monitors
- ApexPro Transmitters
- ApexPro CH Transmitter (not for sale outside of the U.S. and Canada)
- ApexPro FH Transceiver (not for sale outside of the U.S.)
- Apex Oximeter SpO2 Module
- Xpod™ Oximeter
- Accutracker DX Noninvasive Blood Pressure (NBP) Monitor
- DINAMAP® PRO Series Monitors
- Antenna System
- Receiver System
- Unity Network
- CIC Pro Clinical Information Center
- 4 Connection
- 5 Maintenance
- 6 Telemetry Setup
- 7 Admit/View a Patient
Revision D ApexPro Telemetry System 6-9
2001989-134
Telemetry Setup: Telemetry Unit Defaults
Waveforms
The ECG 1 popup list is used to set the unit default for the waveform
that will print in the first position on a graph. Lead II is the default.
The Waveform 2, 3, and 4 options are used to set the unit defaults for the
waveforms that will be printed in the second, third, and fourth positions
on the graph.
The options found in these popup lists are:
I
II
III
V (This option will indicate the V lead being monitored, e.g., V2.)
AVR
AVL
AVF
Off (available for Waveform 2, 3, and 4 only)
Transmitter Graph
The Transmitter Graph option can be set to On or Off. When this option
is set to On, a telemetry patient can initiate a graph by pressing the
Graph button on the telemetry transmitter. When this option is set to
Off, graphs cannot be initiated at the transmitter.
This option sets the unit default for all telemetry patients admitted to
the CIC Pro. When the Transmitter Graph option is set to On,
transmitter graph privileges can also be set on an individual patient
basis using the Enable Transmitter Graph check box in the patient’s
Graph Setup tab sheet.
Alarm Graph
The Alarm Graph On/Off option allows you to select whether a graph
will be printed when an alarm occurs. The choices are:
Always off
Always on (factory default selection)
This option sets the unit default for all patients admitted to the CIC Pro.
It cannot be changed on an individual patient basis.