5-4 Model 340 Telemetry System Revision B
2003720-001
Monitoring via Telemetry: Monitoring Reminders
Tocotransducer
n
Remember to place the receiver’s UA Mode Selector switch in the
TOCO position.
n
When connecting or re-connecting a Corometrics Trimline
Tocotransducer to the transmitter’s
UA connector, you must wait at
least 10 seconds before pressing the monitor’s
UA Reference button. If
any device (monitor, receiver, transmitter is off), you must wait at
least ten seconds from the time the last device is powered on.
IUP
n
Remember to place the receiver’s UA Mode Selector switch in the IUP
position.