Step-by-step Instructions

4-14 Model 340 Telemetry System Revision A
2006899-001
Setup Procedures: Performing a Functional Checkout
Testing the UA Functions
1. Place the receivers UA Mode Selector switch in the TOCO position.
,03257$17
TRIMLINE TOCOTRANSDUCERSIf the monitor is on
when you connect or re-connect a Trimline
Tocotransducer to the
UA connector, you must wait at
least 10 seconds before pressing the
UA Reference button.
If the monitor is off, you must wait at least 10 seconds
from the time the monitor is powered on.
2. Plug a tocotransducer into the transmitters
UA connector. Verify the
following on the monitor:
u
If the monitor has a UA display: the display reads an arbitrary
pressure value.
u
If the monitor does not have a UA display: turn on the strip chart
recorder and check that
TOCO prints on the papers mode
annotation line.
Figure 4-17. Connecting a Tocotransducer or IUPC Cable
,03257$17
DEFAULT REFERENCE VALUEMost monitors have
a default UA reference of 10 relative units. Take into
consideration that newer model monitors can be
configured to store a custom default value.