Step-by-step Instructions

Revision A Model 340 Telemetry System 4-11
2006899-001
Setup Procedures: Performing a Functional Checkout
Testing the Ultrasound Functions
,03257$17
TRANSDUCER TYPEUse only Corometrics 5700
Series Ultrasound Transducers with the Model 340
Telemetry System.
1. Plug an ultrasound transducer into the
Ultrasound connector on the
transmitter.
Figure 4-14. Connecting an Ultrasound Transducer
2. Verify the following:
u
Models 115, 145: The corresponding FHR display reads 0 BPM.
u
Models 116, 118, 150, 151, 118, 155 and Series 120, 170: The
corresponding FHR display shows “– –”.
If the display fails to illuminate, ensure that the corresponding
interconnection cable is firmly attached to both the monitor and the
receiver.
3. Use your finger to gently rub the ultrasound transducer face in a
rhythmic mannerto simulate a FHR. Try to maintain a steady rate
and verify the following on the monitor:
u
the corresponding FHR display value responds to the rubbing;
u
the corresponding FHR heartbeat indicator responds to the
input; and
u
the ultrasound audio tones are synchronous with the transducer
stroking.