Revision A Model 340 Telemetry System 5-7
2006920-001
Calibration: Receiver Calibration
ECG Channel
1. Confirm the 3 dB points of the band-pass frequency are at least
2200 Hz and 3000 Hz between the input TP45 and the output TP17.
Confirm the filter gain of 2 at center frequency.
2. With input frequency set at 2700 Hz and 250 mVp-p at TP45,
adjust R106 until the dc voltage at TP10 is equal to the dc voltage at
TP46 ± 10 mVdc.
3. Sweep the input frequency between 2450 Hz and 2950 Hz at a rate of
approximately 25 Hz. Observe the 25 Hz waveform at TP19 with an
8 ± 1.6 Vpp amplitude.
4. Confirm the PLL lock-in range. This is the frequency where it
first achieves lock, not the range over which it maintains lock.
PLL lock occurs at 2300Hz and3100Hz. Confirm lock occurs at
22052450 Hz and29503245 Hz. Lock-in is verified by a logic low at
TP18.
5. Confirm the frequency response 3 dB points at TP19 are 5 and 50
Hz.
dB
UNITS
AC
VOLTS
888 888 888
Hz Generator
888 888 888
Hz Counter