Operating Instructions
STSRn00 Programmer v.1.0.02
Chan 0 458.5 MHz << Comparison freq. 12.5 KHz
Chan 1 458.5125 MHz Ref. (TCXO) freq. 12. 8 MHz
Chan 2 458.525 MHz Parallel channel select
Chan 3 458.5375 MHz
Chan 4 458.55 MHz
Chan 5 458.5625 MHz
Chan 6 458.575 MHz COMMANDS:
Chan 7 458.5875 MHz
Chan 8 458.6 MHz F2 :copy Ch 16-31 to Ch 0-15
Chan 9 458.6125 MHz F3 :change comparison frequency
Chan 10 458.625 MHz F4 :set parallel ch. mode
Chan 11 458.6375MHz F5 :read from unit
Chan 12 458.65 MHZ F6 :program unit
Chan 13 458.6625MHz F7 :program serial channel
Chan 14 458.675 MHz F12 :QUIT
Chan 15 458.6875MHz
Start table 458.7 MHz
Max. Freq. 458.95 MHz Sequential frequencies, Chan. 16 to 127
Table step 1 x 12.5 kHz
Figure 3
NOTES:
1. The reference (TXCO) frequency is not programmable.
2. The readback function (F5) reads the current settings from the unit into the PC.
3. Function F6 (program the unit) is only enabled after the read-back, (refer to
Figure 2). This function re-programs the unit connected.
4. A value for each parameters has to be entered.
5. Only channel 0 to 15 frequencies can be displayed by this software explicitly.
6. Sequential channel table is displayed as start, step and stop frequencies.
Serial channel selection
The unit defaults to serial channel selection whenever the software is run. Selecting
the F7 function key prompts the user to enter the new serial channel number which is
then displayed in ‘Serial channel selected’.