76 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions Revision A - Draft 7
2001989-135
ApexPro Transmitter
ApexPro Appendices
Appendix 1: ApexPro Troubleshooting
LED Status Problems
Condition Possible Cause Possible Actions
No lights are blinking. Hardware is not connected
properly.
Check the hardware connections as described in “Connect the Hardware” on
page 11.
Software is not installed. Install the Apex & ApexPro Tx Config software as described in “Install
the Programming Device” on page 9.
Software is not running. Start the Apex & ApexPro Tx Config software as described in “Run
the Apex & ApexPro Programming Box Software” on page 13.
Batteries do not have enough
power.
Replace the batteries:
1. Disconnect the transmitter from the programming device.
2. Replace the transmitter batteries with two new AA alkaline batteries.
3. Reconnect the transmitter to the programming device.
One or both firmware files are
corrupt.
Follow the instructions in “Download File” on page 70.
All lights blink rapidly several
times and then slowly two
times.
Normal status when hardware
has just been properly
connected.
No action needed.
All lights blink on and off every
second.
Normal status when hardware
is properly connected and the
software is running.
No action needed.
RA LED light is blinking. You have chosen to “Download
File” as on page 70 and need
to download the .app file.
Follow the instructions in “Erase & Download Application Code” on page 66.
Transmitter is receiving a
download of application
firmware.
No action needed.
LA LED light is blinking. You have chosen to “Download
File” as on page 70 and need
to download the .mfg file.
Follow the instructions in “Erase & Download Service Code” on page 67.
Transmitter is receiving a
download of service firmware.
No action needed.
Draft