User Manual

Table Of Contents
74 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions Revision A - Draft 7
2001989-135
ApexPro Transmitter
a. Select the Change Features button.
b. Click on the desired software level, then select OK. Note that if you
downgrade, you will lose some features.
c. Enter the password. (To get the password, contact Technical
Support. See page 78.) Then select OK.
7. The change is shown on the Program ApexPro Transmitter window.
Check the Optional Feature Status field to verify that the features you
selected were enabled.
Verify Transmitter Operation
Complete the checkout procedures identified in the ApexPro Telemetry
Transmitter Service Manual.
Update Labels and Dust Covers
The screw cover labels and the dust covers are used to visually identify the
software features of the transmitter. If the feature level is changed, the
labels and the dust covers should be updated according to the table below.
520A
530A
Feature Transmitter Appearance
Multi-Lead ECG acquisition with fully-
functional interface connector ports
Grey-colored screw cover label
Grey-colored dust covers
Single Lead ECG acquisition with fully-
functional interface connector ports
Single Lead screw cover label
Grey-colored dust covers
Dust covers
Screw cover labels
Draft