Revision A - Draft 7 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions 71
2001989-135
ApexPro Transmitter
3. A window opens for you to navigate to the I/Q Table you need.
4. Locate the transmitter firmware, generally in C:\Program Files\GE
Medical Systems\Apex and ApexPro Programming Box Software, or
where the firmware was installed.
5. Click once on the .tbl file and select Open.
6. Select OK.
7. The transmitter’s settings will disappear temporarily from the Program
ApexPro Transmitter window. When the settings reappear, verify that
the I/Q Table Version matches the last three digits of the .tbl file name
listed on the CD.
NOTE
NOTENOTE
NOTE
The version has 4 digits and is stored inside the table itself. The version
number may not correlate to the file name or to the firmware part
number.
8. Confirm that all Power-up Self Test Status messages indicate Passed. See
“Appendix 1: ApexPro Troubleshooting” on page 76 if any failures occur.
445A
450A
455A
Draft