- 3 ApexPro CH Transmitter
- Program the Transmitter’s Basic Functions
- View Transmitter Diagnostics
- Update Transmitter Firmware
- ApexPro CH Appendices
- 4 ApexPro Transmitter
Revision A - Draft 7 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions 69
2001989-135
ApexPro Transmitter
6. Click once on the .app file and select Open. The upgrade process starts
automatically.
7. During the upgrade, the message box on the Program ApexPro
Transmitter window displays a countdown timer and the transmitter
settings appear greyed out. In addition, sometimes the RA LED light on
the transmitter flashes and sometimes all the LED lights flash.
8. Next, the download process downloads the .mfg file. Select OK at this
screen.
9. Locate the transmitter firmware, generally in C:\Program Files\GE
Medical Systems\Apex and ApexPro Programming Box Software, or
where the firmware was installed.
10. Click once on the .mfg file and select Open. The upgrade process starts
automatically.
11. During the upgrade, the message box on the Program ApexPro
Transmitter window displays a countdown timer and the transmitter
410A
415A
420A
425A