User Manual

Table Of Contents
Revision A - Draft 7 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions 67
2001989-135
ApexPro Transmitter
4. Select OK at this screen.
5. Locate the transmitter firmware, generally in C:\Program Files\GE
Medical Systems\Apex and ApexPro Programming Box Software, or
where the firmware was installed.
6. Click once on the .app file and select Open. The upgrade process starts
automatically.
7. During the upgrade, the message box on the Program ApexPro
Transmitter window displays a countdown timer and the transmitter
settings appear greyed out. In addition, sometimes the RA LED light on
the transmitter flashes and sometimes all the LED lights flash.
8. Wait until the first line in the message box displays the text Connected to
the transmitter. If there are any problems, refer to “ApexPro Appendices”
on page 76.
Erase & Download Service Code
1. Access the high-level technical functions of the software as described in
“Access the Super User Mode” on page 79.
2. The Program ApexPro Transmitter window will display additional
features, including a Download Functions section (reference D in
“Appendix 2: Access ApexPro Technical Functions” on page 79.)
3. Select the Erase & Download Mfg button.
4. Select OK at this screen.
405A
410A
415A
420A
Draft