Revision A - Draft 7 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions 37
2001989-135
ApexPro CH Transmitter
Reference Lead Setting
For 5- or 6-Lead Cables
The reference lead setting has no effect when 5- or 6-lead cables are used.
For 5- or 6-lead cables, the reference lead defaults to RL.
For 3-Lead Cables
For 3-lead cables, one of the lead wires (LL, LA, or RA) is used to connect the
reference voltage to the patient. Selection of the reference lead determines
which ECG waveform will display at the CIC.
Select the desired Reference Lead.
Alarm Pause Setting
The purpose of the alarm pause setting is to allow clinicians to adjust the
transmitter and/or the patient without setting off alarms. The alarm pause
setting is an amount of time, in minutes, during which alarms will not
activate. After this time has elapsed, the alarm pause will automatically be
deactivated and the transmitter will resume normal operation. You can
manually turn off the alarm pause at any time by pressing Verify Leads
and Graph at the same time.
NOTE
NOTENOTE
NOTE
Activation and deactivation of the alarm pause is done on the
transmitter itself. The setting in the software only sets the number of
minutes to wait until the alarm pause times out.
Select this
reference lead...
to view this
waveform at the CIC
Lead I LL (left leg)
Lead II LA (left arm)
Lead III RA (right arm)
315A