Test Status Failures
Technical Support
If the above information does not resolve your problem, call:
Condition Possible Cause Possible Actions
EEPROM Checksum indicates
fail status.
The transmitter firmware is
incorrect or corrupt.
Cycle the power on the transmitter. If the failure occurs again, re-program
the transmitter’s functions and cycle the power on the transmitter. If the
failure still occurs, contact Technical Support. (See below.)
Synthesizer Lock indicates fail
status.
The transmitter is currently
unable to operate at its
programmed frequency.
Try programming the transmitter to another frequency. If the failure still
occurs, contact Technical Support. (See below.)
U.S. and Canada 800-558-7044