User Manual

Table Of Contents
Revision A - Draft 7 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions 13
2001989-135
Introduction
Run the Apex & ApexPro Programming Box Software
Start the Software
1. Complete the procedure “Install the Programming Device” on page 9.
2. Complete the procedure “Connect the Hardware” on page 11.
3. Select Start > Programs > GE Medical Systems > Apex & ApexPro Tx
Config > Apex & ApexPro Tx Config to start the programming software.
You may also have a shortcut on the desktop called Apex & ApexPro Tx
Config to start the software.
Select the Transmitter
Select the type of transmitter to be programmed and the communications
port, as described below.
1. Select the type of transmitter.
2. Select the correct COM Port.
If you know which COM Port is being used, select that. COM1 is most
common.
If you do not know the correct COM Port, select COM1 and then OK. If
this is not the correct COM Port, an error message No communication
from the transmitter will appear in the message area of the Program
ApexPro Transmitter window. You can then try COM2 and so on. See
“The Main Programming Window” on page 23 for explanation of that
window.
3. Select OK. The software remembers the settings.
Table 1. Select Transmitter
Ref Definition
A Type of transmitter to be used
B COM port on the computer to which the programming box cable is connected
A
105A
B
Draft