User Manual Part 7

D-8 ApexPro™ 2001989-301A- draft 1
Troubleshooting
ALARM PAUSE All alarms for this patient have been turned off for five
minutes. This is initiated from the transmitter by pressing
both the Verify Leads and Graph buttons simultaneously.
Refer to ALL ALARMS OFF.
ALL ALARMS OFF All alarms for this patient have been turned off. No graph
strips run, arrhythmia events are not stored, and no audible
tones sound if an alarm condition should occur.
ARR OFF The arrhythmia program for a selected patient has been
turned off. No arrhythmia messages are displayed,
arrhythmia event s are not stored, no graph strips run, and no
audible tones sound if an arrhythmia alarm condition occurs.
ARTIFACT ARR SUSPEND All artifact begins at level 1. Sustained artifact progresses to
level 2 when noise on ECG lasts for 20 of the last 30
seconds.
Level 1 — Upon immediate detection of artifact, the
message ARTIFACT is displayed. There is no alarm
tone. Only lethal arrhythmia processing is available.
Level 2 — Arrhythmia monitoring is suspended in this
condition. A system warning alarm sounds, no
arrhythmia messages are displayed, no graph strips run
and no arrhythmia events are stored. Heart rate and
PVC values change to X, an additional message, ARR
SUSPEND, is displayed.
LEADS FAIL All leads have failed, right leg lead failed, lead wires
unplugged or reference lead failed. If set to Warning or
Advisory level, a system alarm is heard. This will self-
silence if condition is corrected, or the user can silence for
one minute with the Silence Alarms button at the bottom of
the CIC Pro center display.
Draft