User Manual Part 7

Customized defaults worksheet
2001989-301A- draft 1 ApexPro™ B-3
Telemetry parameter limits and alarm levels
Patient Age Adult
Transmitter Pause Enabled
Alarm Pause Breakthrough Always On
Event Marker
1
On
1
The
Event Marker Graph
and
Event Marker
features are not applicable to all transmitters.
Telemetry unit default Default setting
Parameter Limits and Alarm Levels Low High Level Arrhythmia Alarm Levels Levels
HR bpm 50 150 Warning Asystole Crisis
ST-I mm -2.0 2.0 Warning VFIB/VTAC Crisis
ST-II mm -2.0 2.0 Warning V Tach Crisis
ST-III mm -2.0 2.0 Warning VT > 2 Crisis
ST-V mm -2.0 2.0 Warning V Brady Crisis
ST-V2 mm -2.0 2.0 Warning Acc Vent Advisory
ST-V3 mm -2.0 2.0 Warning Pause Advisory
ST-V4 mm -2.0 2.0 Warning Tachy Advisory
ST-V5 mm -2.0 2.0 Warning Brady Advisory
ST-V6 mm -2.0 2.0 Warning R on T Message
ST-AVR mm -2.0 2.0 Warning Couplet Message
ST-AVL mm -2.0 2.0 Warning Bigeminy Message
ST-AVF mm -2.0 2.0 Warning Trigeminy Message
NBP-S mmHg 80 200 Warning PVC Message
NBP-D mmHg 20 120 Warning Irregular Message
NBP-M mmHg 40 140 Warning Atrial Fib Message
SPO2 % 90 105 Warning
SPO2-R bpm 50 150 Warning System Alarm Levels Levels
RR breaths/min 5 30 Warning CHANGE BATTERY System Warning
RR-APNEA seconds 30 Warning OFF NETWORK System Warning
PVC #/min 6 Advisory ARR SUSPEND System Warning
LEADS FAIL System Warning
PROBE OFF System Warning
Draft