User Manual Part 6

4-16 ApexPro™ 2001989-301A- draft 1
System setup
Current telemetry listings
NOTE
Refer to the CIC Pro Clinical Information Center Bedrock Platform Service
Manual for important configuration information.
Duration Designate how much data is included in the strip. The
maximum strip duration is 60 minutes.
To set the strip duration, place the cursor on the scroll bar
below the Strip Duration display field. Move the scroll bar to
the left for shorter duration or to the right for longer duration.
Hole Location Provide space for binding printed reports. Choices are:
none, top, bottom, left, and right.
Unit License Default: Full
Disclosure License Type
NOTE
You must be in the Service mode at the CIC Pro center
to modify this setting.
Display a list of the full disclosure license type. Choices are:
none, 24 hours, 48 hours, and 72 hours.
NOTE
If the default does not match the actual license, full
disclosure does not work.
Offline Storage Select a time period to store full disclosure data if contact
with a monitor has been lost. Choices are: 30mins, 1, 2, 4, 8
and 12 hours.
WARNING
POTENTIAL DATA LOSS — Do not allow a NO COMM
(patient offline) event to exceed the time limit selected in the
Offline Storage setting. The patient’s full disclosure data is
deleted if the time limit is exceeded.
For more information, refer to the CIC Pro Clinical
Information Center Bedrock Platform Service Manual.
Start Data Storage NOTE
You must be in the Service mode at the CIC Pro center
to modify this setting.
Designate how full disclosure is enabled for patients at the
time of admission. Choices are: automatically for all beds,
automatically if listed, and manually.
Bed List NOTE
You must be in the Service mode at the CIC Pro center
to modify this setting.
Lists beds for which full disclosure data is automatically
stored.
Restore Clear any changes you made to the full disclosure default
settings and revert to the previous settings.
Option Function
Draft