User Manual Part 6

System setup
2001989-301A- draft 1 ApexPro™ 4-13
To view the telemetry unit default settings, click CIC Setup > Telemetry Alarm
Control Defaults.
501A
Parameter Limits and Alarm Levels Low High Level
HR bpm 50 150 Warning
ST-I mm -2.0 2.0 Warning
ST-II mm -2.0 2.0 Warning
ST-III mm -2.0 2.0 Warning
ST-V mm -2.0 2.0 Warning
ST-V2 mm -2.0 2.0 Warning
ST-V3 mm -2.0 2.0 Warning
ST-V4 mm -2.0 2.0 Warning
ST-V5 mm -2.0 2.0 Warning
ST-V6 mm -2.0 2.0 Warning
ST-aVR mm -2.0 2.0 Warning
ST-aVL mm -2.0 2.0 Warning
ST-aVF mm -2.0 2.0 Warning
NBP-S mmHg 80 200 Warning
NBP-D mmHg 20 120 Warning
NBP-M mmHg 40 140 Warning
SPO2 % 90 105 Warning
SPO2-R bpm 50 150 Warning
Draft