User Manual Part 6
4-12 ApexPro™ 2001989-301A- draft 1
System setup
Factory defaults
These factory defaults are in effect, depending upon the patient’s age, unless they
have been modified through Telemetry Unit Defaults.
ECG Display Lead is II
Multi-Lead analysis
Heart rate alarm limits (high/low):
Adult—150/50
0–2 years—200/90
3–10 years—180/60
11–13 years—150/50
ST measurement:
Adult— J+ 60ms
0–2 years— J+ 30ms
3–10 years— J+ 40ms
11–13 years— J+ 50ms
PVC limit is 6
1X size
Pace off
Arrhythmia On (arrhythmia changes with age)
ST off
Graph leads II and V
25 millimeters per second speed
Alarm Graph location at the CIC Pro center where patient was admitted
TTX (manual) graph location at the CIC Pro center where patient was admitted
Print window location at the CIC Pro center where patient was admitted
Telemetry alarm control defaults
This option sets the telemetry default alarm limits and alarm level settings. In user
mode, all of the controls are view-only. You must be in Service mode to set the
Telemetry Alarm Control Defaults at the CIC Pro center.
3-10 years Adult
0-2 years
3-10 years
11-13 years
50, 150
90, 200
60, 180
50, 150
Message
Message
Message
Message
Advisory
Crisis
Message
Advisory
11-13 years Adult
0-2 years
3-10 years
11-13 years
50, 150
90, 200
60, 180
50, 150
Message
Message
Message
Message
Advisory
Crisis
Crisis
Message
Telemetry Unit Defaults
Tab Sheet Setting
Age Chosen In The Admit Tab
Sheet When Patient Admitted
Resulting Limits
Unit Default
Alarm Level (Brady)
Resulting
Alarm Level
Draft