User Manual Part 6
5-12 ApexPro™ 2001989-301A- draft 1
Alarms
2. To turn off alarms for this telemetry patient, click OFF.
WARNING
The telemetry alarms remain off until you manually select ON again.
Telemetry bed alarm control defaults
You cannot adjust the Telemetry Alarm Control Defaults. These settings are service-
level defaults and are password protected.
To view the settings, from the multi-patient viewer, click CIC Setup > Telemetry
Alarm Control Defaults.
To permanently change the Telemetry Alarm Control Defaults, see the CIC Pro
Clinical Information Center Bedrock Hardware Platform Service Manual.
Alarm unit default settings
Telemetry alarm control defaults
You can set Telemetry Unit Defaults for Parameter Limits And Alarm Levels, as
well as for Arrhythmia Alarm Levels. These defaults are in effect for all telemetry
patients admitted to your unit, unless they are modified in an individual patient’s
Alarm Control tab sheet.
To set Telemetry Alarm Control Defaults, follow the steps below:
1. Click CIC Setup. A set of tabs displays.
2. Click on the Service Password tab to bring it to the front.
Cardiac rehab CAR REHAB CARDIAC REHAB
GI Lab GI LAB GI LAB
Occupational therapy O.T. OCCUPATIONAL THERAPY
Off unit OFF UNIT OFF UNT
Cath Lab CATH LAB CARDIAC CATH LAB
1
OFF appears dimmed and is not selectable when the following service-level default is set:
CIC
Setup
>
CIC Defaults
>
Allow Alarms OFF on this CIC
>
No
. To be able to turn off alarms
for a telemetry patient, you must change the
Allow Alarms OFF
on this CIC service-level
default to
Yes
. For more information, refer to the CIC Pro Clinical Information Center
Bedrock Hardware Platform Service Manual.
Alarms Off Reason Text Displayed At CIC Pro center Text Printed On Graph