Note: Once a month, or whenever the DT-4500 is submersed or subjected to a
stream of liquid, remove the accessory connector cover and remove any
moisture that may have collected inside.
Cleaning the Battery Compartment
CAUTION: When cleaning the battery compartment, use only soap, water, or alcohol. Do
not use any other cleaning agents inside the battery compartment as they
may damage the battery compartment.
CAUTION: Make sure to rinse all cleaned surfaces with distilled water to remove any
cleaning agent residue. Dry off the battery contact leads. Ensure that the bat-
tery compartment is dry before inserting the battery into the unit.
Under normal operation, the battery compartment should not require frequent clean-
ing. If the battery compartment does require cleaning, then use the following instruc-
tions.
1. Remove the battery from the battery compartment.
2. Clean the transceiver with a gauze pad or cloth moistened with one of the fol-
lowing agents:
Water
•Soap
3. Use a cloth moistened with distilled water to rinse away the cleaning solution.
4. Dry thoroughly with a lint-free cloth. Allow the battery compartment to air
dry completely prior to closing the compartment door.
Cleaning the ECG Leadsets
The transceiver ECG Leadsets are manufactured by Affinity Medical.
Contact your technical support representative for additional leadsets.
Warnings
1. Do not use leadsets which exhibit signs of wear or damage such as crack-
ing or degradation of the connectors or cable insulation.
2. Do not sterilize using steam or gamma radiation. Damage to the leadsets
will result.
Cautions
1. To increase the life of the leadsets, do not pull on the leadsets to discon-
nect. Pull gently by grasping the connectors.
2. Do not immerse the leadsets in water or other liquid to clean. Immersion
may cause damage to the leadsets.
3. Repeated exposure to EtO sterilization will shorten the effective life of the
leadset. The leadsets should be sterilized only when indicated by specific
patient or hospital requirements.