User's Manual

WMTS TRANSCEIVERS
PatientNet Operator’s Manual, v1.04, 10001001-003, Revision B 193
All information contained herein is subject to the rights and restrictions on the title page.
Cleaning
This section provides cleaning and maintenance instructions for DT-4500
transceivers.
Read and follow all precautions when cleaning transceivers.
WARNING: No claims are made concerning the sterility of the DT-4500 Ambulatory Trans-
ceivers.
CAUTION: Do not sterilize any part of the transceivers. Gas sterilization, autoclaving, liq-
uid immersion, and other sterilization methods can cause serious damage to
the devices that may not be obvious to the user.
Note: DO NOT use abrasive cleaners.
Cleaning the Chassis
The following applies to cleaning the DT-4500.
The DT-4500 can be cleaned with the patient cable attached, however please
ensure that the cleaning agents used to clean the DT-4500 are compatible with
the cleaning agents listed for the ECG cable on page 194, or else ensure that
the ECG cable does not come into contact with the cleaning agents for the DT-
4500.
To clean around the ECG connector, remove the ECG cable from the unit.
CAUTION: Prior to cleaning the battery compartment and transceiver chassis, remove
the battery from the unit.
1. Remove the battery from the transceiver and inspect the battery compartment
after each use. Close the battery door.
CAUTION: Prior to rinsing the DT-4500, make sure that the battery compartment door is
properly closed and sealed.
2. Transceivers can be cleaned with a gauze pad or cloth moistened with one of
the following agents:.
Soap and Water
Quaternary Ammonium
Glutaraldehyde 2%
Dilute Chlorine Bleach (Sodium hypochlorite), 10% solution, freshly
made in past 24 hours
Isopropyl Alcohol 70%
Ethyl Alcohol
3. Use a cloth moistened with distilled water to rinse away the cleaning solution.
4. Dry thoroughly with a lint-free cloth.