User manual

R
Pulse Counting
Pulse Count with GAMMA-SCOUT
©
• PressthebuttontoswitchGAMMA-SCOUT
®
to the
pulse count mode. The display will show the pulse symbol.
Counting will not have started yet. Pressing the button
a second time will start the counting without pre-setting the
measuring time.
• Pressthebuttonifyouwanttosetameasuringtime:
For the measuring time to count in seconds, press
the button once.
For the measuring time to count in minutes, press
the button twice.
For measuring time to count in hours, press the
button three times.
• Nowsettheexactvalueofthedesiredmeasuringtime
using the and buttons.
• Startmeasuringbypressingthebuttonasecond
time.Thepulsesymbolashesinthedisplayduringthe
measuring time.
• Ifameasuringtimewasset,thepulsesymbolashes
until the end of the measuring time, after which it is steady
again. The display now shows the number of pulses coun-
ted for the programmed interval.
• Youcanstopmeasuringintwoways:
By pressing the button once again, the measuring
result remains visible in the display.
By choosing another mode of operation, the measuring
result disappears.
When and Why?
GAMMA-SCOUT
®
can also be used as a normal Geiger counter. In this case, it simply counts
the number of pulses received without converting them into the sievert unit.
Since the conversion is dependent upon the isotope of the radiation source, and since we based
our conversion on Cs-137, this option is important for users who work with other isotopes and want
to convert using personal charts.
GAMMA-SCOUT
®
stores the received pulses in its internal memory.
7