Operation Manual

Physical Parameter Si unit Old unit Relation
Activity Becquerel (Bq) Curie (Ci) 1 Ci = 3.7 * 10
10
Bq
1 Bq = 1/s 1 Bq = 2.7 * 10
-11
Ci
= 27 pCi
Ion dose I Coulomb / kg Röntgen (R) 1 R = 2.58 * 10
-4
C/kg
1 C/kg = 3876 R
Energy dose D Gray (Gy) Rad (rd) 1 rd = 0.01 Gy
1 Gy = 100 rd
Equivalent dose H sievert (Sv) Rem (rem) 1 rem = 0.01 Sv
1 Sv = 100 rem
Effective Dose H
E
sievert (Sv) Calculated value /
1 Sv = 1 J / kg radiation protection
For relationship between Becquerel and sievert see pages 20-21.
For further information on Becquerel, see Gamma FAQ on www.gamma-scout.com
Calibration
Under environmental radiation, the counter tube is not subject to fatigue and, therefore, will not re-
quire re-calibration. However, if the user holds ISO certication, periodical calibration is mandatory.
For this purpose, we offer the following service:
Testing is sub-contracted to an assembly operation, which tests it for 72 hours against a master.
The master is calibrated against a gauged reference source (Cs-137). A data log is then generated.
To the best of our knowledge, this record is accepted as compliant by ISO auditors without excepti-
on.
Currently, this costs 50 Euro before sales taxes.
IT-Support
In order of problems by installing the product drivers or TOOLBOX software, do not hesitate to
contact us:
GAMMA-SCOUT GmbH & Co. KG
Abtsweg 15
D-69198 Schriesheim
Germany
Fax: +49 (0) 6220 / 6640
E-Mail: e.mirow@gamma-scout.com
R
Technical Data, Calibration