User's Guide
Dexcom G6 System User Guide
303
Appendix F: Technical Information
Table 4-B. Trend Accuracy Rate of Change (Pediatrics*; n=165)
CGM Rate
Range
(mg/dL/min)
YSI glucose range (mg/dL)
CGM-YSI
Pairs
(n)
<-2 [-2,-1) [-1,-0) [0,1] (1,2] >2
<-2 47.9% 37.0% 12.8% 1.9% 0.0% 0.5% 211
[-2,-1) 6.6% 55.5% 33.8% 3.4% 0.6% 0.1% 686
[-1,0) 0.5% 8.9% 73.7% 15.8% 1.0% 0.0% 2,048
[0,1] 0.0% 0.8% 25.5% 62.9% 10.0% 0.8% 1,666
(1,2] 0.0% 0.4% 4.4% 35.9% 48.0% 11.4% 546
>2 0.0% 0.5% 1.7% 7.1% 23.6% 67.1% 423
* Includes pediatric subjects 6-17 years of age; no YSI measurements were taken for
pediatric subjects 2-5 years of age.
LBL014003 Rev XXX Artwork, G6 CGM System User Guide US.indd 303 2/5/19 9:27 AM