User's Guide

Dexcom G6 System User Guide
323
Appendix F: Technical Information
Electromagnetic Immunity and Emissions Declaration
and Guidance
The transmitter and receiver are intended for use in the electromagnetic environment
specified in the next table. The customer or the user of the transmitter should ensure
that it is used in such an environment.
Immunity Test
Transmitter Compliance
Level
Receiver Compliance Level
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
Magnetic Field (50Hz)
IEC 61000-4-8
30 A/m
Electrical Fast
Transient/Burst
IEC 61000-4-4
N/A ± 2 kV for power supply lines
Surge
IEC 61000-4-5
N/A ± 1 kV line(s) to line(s)
Voltage Dips and
Interruptions
IEC 61000-4-11
IEC 60601-1-11
N/A
0% 230V for 1 cycle
0% 230V for 0.5 cycle at 8
phase angles
70% 230V (30% dip in 230V)
for 25 cycles
0% 230V for 250 cycles
Conducted Fields
Disturbance
IEC 61000-4-6
N/A
6 Vrms
150 kHz to 80 MHz
Radiated Fields
Disturbance
IEC 61000-4-3
10 V/m at 80 MHz to 2700 MHz (AM Modulation)