User's Guide
Dexcom G6 System User Guide
298
Appendix F: Technical Information
Table 3-A. Concurrence of G6 CGM Readings and YSI Values by CGM Glucose
Range (Adults; n=159)
CGM
Glucose
Range
1
(mg/dL)
YSI (mg/dL)
< 40
40-
60
61-
80
81-
120
121-
160
161-
200
201-
250
251-
300
301-
350
351-
400
>
400
Total
<40 13.5% 56.7% 24.0% 3.8% 1.9% . . . . . . 104
40- 60 1.2% 67.8% 27.9% 2.7% 0.2% 0.1% . . . . . 917
61- 80 0.1% 21.3% 61.4% 16.9% 0.3% 0.1% . . . . . 2,275
81- 120 . 0.4% 13.6% 70.3% 15.1% 0.6% 0.0% . . . . 3,782
121- 160 . . 0.0% 14.2% 64.3% 20.1% 1.3% 0.0% 0.0% . . 3,026
161- 200 . . . 0.1% 14.5% 56.7% 26.9% 1.5% 0.2% 0.0% . 2,597
201- 250 . . . . 0.2% 12.1% 59.4% 25.4% 2.9% 0.0% . 2,869
251- 300 . . . . . 0.1% 13.7% 59.1% 25.3% 1.9% . 2,268
301- 350 . . . . . . 0.2% 22.3% 63.4% 13.7% 0.5% 1,212
351- 400 . . . . . . . 0.8% 43.9% 52.5% 2.9% 383
>400 . . . . . . . . 5.9% 76.5% 17.6% 34
1
CGM readings are within 40 to 400 mg/dL, inclusive.
LBL014003 Rev XXX Artwork, G6 CGM System User Guide US.indd 298 2/5/19 9:27 AM