User's Guide

Dexcom G6 System User Guide
20
Chapter 2: Indications for Use and Safety Statements
Safety Statements in Chapters
Each chapter shows applicable indications, contraindications, precautions, and
warnings. Some chapters have multiple Safety Statements; others have none. The
same statement may be repeated in several chapters.
2.3 Dexcom G6 Safety Statements
Indications for Use
The Dexcom G6 Continuous Glucose Monitoring System (Dexcom G6 System) is a real
time, continuous glucose monitoring device indicated for the management of diabetes
in persons age 2 years and older.
The Dexcom G6 System is intended to replace fingerstick blood glucose testing for
diabetes treatment decisions. Interpretation of the Dexcom G6 System results should
be based on the glucose trends and several sequential readings over time. The Dexcom
G6 System also aids in the detection of episodes of hyperglycemia and hypoglycemia,
facilitating both acute and long-term therapy adjustments.
The Dexcom G6 System is also intended to autonomously communicate with digitally
connected devices, including automated insulin dosing (AID) systems. The Dexcom
G6 System can be used alone or in conjunction with these digitally connected medical
devices for the purpose of managing diabetes.
Contraindication
No MRI/CT/Diathermy – MR Unsafe
Don’t wear your CGM (sensor, transmitter, receiver, or smart device) for magnetic
resonance imaging (MRI), computed tomography (CT) scan, or high-frequency
electrical heat (diathermy) treatment.
The G6 hasn’t been tested in those situations. The magnetic fields and heat could
damage the components of the G6, which may cause it to display inaccurate G6
sensor glucose readings (G6 readings) or may prevent alerts. Without G6 readings
or alarm/alert notifications, you might miss a severe low or high glucose event.
MR
LBL014003 Rev XXX Artwork, G6 CGM System User Guide US.indd 20 2/5/19 9:26 AM