OPERATOR'S MANUAL COLOR LCD SOUNDER MODEL FCV-1150 www.furuno.
The paper used in this manual is elemental chlorine free. ・FURUNO Authorized Distributor/Dealer 9-52 Ashihara-cho, Nishinomiya, 662-8580, JAPAN Telephone : +81-(0)798-65-2111 Fax : +81-(0)798-65-4200 All rights reserved. Printed in Japan A : DEC . 2007 C : NOV . 16, 2011 Pub. No.
IMPORTANT NOTICES General • The operator of this equipment must read and follow the descriptions in this manual. Wrong operation or maintenance can cancel the warranty or cause injury. • Do not copy any part of this manual without written permission from FURUNO. • If this manual is lost or worn, contact your dealer about replacement. • The contents of this manual and equipment specifications can change without notice.
SAFETY INSTRUCTIONS The operator and installer must read the applicable safety instructions before attempting to install or operate the equipment. WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, can result in minor or moderate injury. Warning, Caution Prohibitive Action Mandatory Action WARNING WARNING ELECTRICAL SHOCK HAZARD Do not open the equipment.
SAFETY INSTRUCTIONS CAUTION CAUTION Do not transmit with the transducer out of water. Properly adjust the gain. Too little gain may present no picture. Too much gain shows excessive noise on the picture. Using the depth data for navigation when the gain is incorrectly set can lead to a dangerous situation. The transducer will be damaged. Do not use depth information for navigation while the picture advance speed is set to "stop".
TABLE OF CONTENTS FOREWORD ........................................ v SYSTEM CONFIGURATION .............. vi 1. 1.20.3 1.20.4 1.20.5 1.20.6 OPERATION ................................. 1 1.1 Control Description......................1 1.2 Power On/Off ..............................2 1.3 Adjusting Display Brilliance .........2 1.4 Selecting a Display......................2 1.4.1 Single frequency display......2 1.4.2 Dual frequency display ........3 1.4.3 Zoom display .......................3 1.4.
FOREWORD A Word to FCV-1150 Owners • Automatic operation selects correct range and gain to show fish echoes and bottom in both shallow and deep waters. Congratulations on your choice of the FURUNO FCV-1150 Color LCD Sounder. We are confident you will see why the FURUNO name has become synonymous with quality and reliability. • In addition to the conventional gain setting, Gain may be applied to the whole display. • Alarms: Bottom, Fish (bottom lock and normal), Speed*, Water Temperature*, and Arrival*.
SYSTEM CONFIGURATION DISPLAY UNIT CV-1150 NavNet 3D Navigation Equipment Power 12-24 VDC Rectifier PR-62 Satellite Compass 100/110/115/ 220/230 VAC 1φ, 50/60Hz Water Temp. Sensor T-02MSB, etc. Water Temp.+Speed Sensor ST-02MSB, etc. : Standard supply : Optional supply : Local supply, external equipment High Freq. Low Freq.
1. OPERATION 1.1 Control Description HOW TO REMOVE THE HARD COVER While pressing the center of the cover with your thumbs as illustrated, pull the cover towards you. 1 2 LF DUAL HF ZOOM ZOOM SHIFT 3 4 PUSH AUTO PUSH 5 6 5 PUSH AUTO 7 5 8 MENU ESC USER-1 ENTER USER-2 0 RANGE MODE LF FUNCTION 0 10 GAIN 10 HF MARK BRILL 10 9 No. Control Function 1 S SHIFT T Shift the display range. (Inoperative when auto range or auto shift is active.) 2 MODE Select display mode.
1. OPERATION 1.2 Power On/Off 1.4 Selecting a Display The FCV-1150 has seven display modes: low frequency display, high frequency display, dual frequency display, low frequency+zoom, high frequency+zoom, user 1 display, and user 2 display. Select one with the MODE control as shown in the procedure below. Press the /BRILL key to turn on the power. The unit beeps, the startup screen appears and then the last-used display is shown. To turn off the power, press the /BRILL key more than three seconds.
1. OPERATION 1.4.3 Gain - Left: Low freq. Picture advance speed - Right: High freq. Output power lowered Range MANUAL L:2.0/H:2.0 LF * Zero line Data box Fish echo Color bar 1/2 P/R AUTO Alarm icon 0 0.0 11.1kt SOG 8.2°C Zoom mode expands chosen area of the single frequency picture. Five modes are available: bottom lock, bottom zoom, marker zoom and two bottom discrimination modes. The default mode is bottom lock. You can preset the zoom mode to use on the Sounder menu.
1. OPERATION Bottom zoom display Bottom discrimination 1/2 display This mode expands bottom and bottom fish on the left-half window, and is useful for determining bottom contour. When the bottom depth increases, the display automatically shifts to keep the bottom echo at the lower part of the screen. The bottom discrimination 1/2 screen shows the single picture on the right half of the screen and the bottom discrimination display occupies the left half of the screen.
1. OPERATION 1.4.4 User 1 and 2 displays The user displays let you customize displays as desired. Two displays are provided and the default settings for each are User 1 display: This screen is split vertically three ways and is comprised of LF, HF and MIX displays. 0 0 20 20 0.0 1. Push the RANGE control to open the auto range setting window. 20 40 49.6 m 60 60 60 Low frequency display High frequency display 2. Use S or T to select Off or On as appropriate.
1. OPERATION 1.6 2. Rotate the RANGE control to select a range. The default ranges are as shown in the table. Unit The basic range and range shifting functions used together give you the means to select the depth you can see on the screen. The basic range can be thought of as providing a "window" into the water column and range shifting as moving the "window" to the desired depth.
1. OPERATION 1.7 2. Use S or T to select Fishing, Cruising or Off as appropriate. Fishing: This mode clearly displays weaker echoes and is for searching fish schools. "H (L):AF" is shown at the top left corner on the screen. Crusing: This mode clearly displays stronger echoes (for example, bottom) and suppresses weak echoes. Use it for general cruising. "H (L):AC" is shown at the top left corner on the screen. Off: Select to manually adjust gain.
1. OPERATION 1.8 Measuring Depth 2. Use S or T to select a menu. The cursor (yellow) highlights current selection. The items in the right window change with menu selected. The VRM (Variable Range Marker) measures the depth to fish schools, etc. 3. Press the ENTER key to send the cursor to the menu item window. (Alternatively, you can press X.
1. OPERATION 1.10 3. Use S to select Pic. Advance and press the ENTER key. Picture Advance Speed The picture advance speed determines how quickly the vertical scan lines run across the screen. When choosing a picture advance speed, keep in mind that a fast advance speed will expand echoes horizontally on the screen and a slow advance speed will contract them. A fast advance speed is useful for observing the rugged bottom closely. A slow advance speed is useful for observing the smooth bottom.
1. OPERATION 1. Press the MENU/ESC key to open the menu. 2. Use S or T to select Sounder and press the ENTER key. 2. Use S or T to select Sounder and press the ENTER key. 3. Use S or T to select Color Erase and press the ENTER key. 3. Use S or T to select Interference and press the ENTER key. 4. Use S or T to turn the interference rejector on or off and press the ENTER key. Off: Turn off the interference rejector.
1. OPERATION 3. Use S or T to select Clutter and press the ENTER key. 7. Use S or T to select setting desired and press the ENTER key. 8. Press the MENU/ESC key several times to close the window. 1.14 Adjusting TVG A fish school at a deep depth is displayed in weak colors even if it is equal in strength to one in shallow waters. This is due to propagation attenuation of the ultrasonic wave. To compensate for this difference, use TVG.
1. OPERATION 3. Use S or T to select TVG and press the ENTER key. 1.15 A-scope Display This display shows echoes at each transmission with amplitudes and tone proportional to their intensities, on the right side of the screen. It is useful for estimating the type of fish school and bottom composition. Note: In the horizontal split dual frequency display the a-scope display appears on both high and low frequency displays.
1. OPERATION 5. Press the MENU/ESC key twice to close the window. MANUAL L:2.0/H:2.0 LF Fish (normal) alarm: The fish (normal) alarm tells you when an echo above a certain strength (selectable) is within the preset alarm range or an echo of yellow color or stronger color is anywhere between the transducer and the bottom. 1/2 0 Weak reflection (small fish or noise) 20 Fish (bottom lock) alarm: The fish (bottom lock) alarm sounds when fish are within a certain distance from the bottom.
1. OPERATION 3. Use S or T to select an alarm and press the ENTER key. Bottom and Fish (B/L) alarms 10. Use S to select Span (or Radius in case of Arrival) and press the ENTER key. 11. Use S or T to set the range of depth, temperature, speed or distance as appropriate. To shorten the alarm range marker, use S and to lengthen it use T. Temperature, Speed and Arrival alarms Fish (Normal) alarm 12.
1. OPERATION 1.17 FUNCTION Control 1.17.2 The FUNCTION control provides for instant display of a user-defined options window, chosen with "FUNC Key" on the Key menu. 15 items are available: Pic. Advance, Interference, Color Erase, Clutter, White Line, White Marker, TVG (default setting), STC, Bottom Zone, A-Scope, Split Range, Auto Shift, Freq Choice, Freq Control and Gain/Mode. Programming the FUNCTION control 1. Push and hold down the FUNCTION control until the FUNC key setting window appears.
1. OPERATION 1.18 Waypoints quential waypoint number. If you want to change the waypoint name, go to step 4. Otherwise, go to step 6. Waypoints are used to: • Record the position of an important echo as a waypoint, and 20 points may be saved. • Output a waypoint position to a chart plotter to mark position on its screen. • Find range, bearing and time-to-go to a location (waypoint). This feature requires position data, fed from a GPS navigator. 1.18.
1. OPERATION Entering a waypoint by manual entry of position 6. Enter latitude and longitude, similar to how entered waypoint name. 1. Press the MENU/ESC key to open the menu. 7. Press the MENU/ESC key to register the waypoint. 2. Use S or T to select Data and press the ENTER key. 8. Press the MENU/ESC key several times to close the window. 1.18.2 Editing waypoints 1. Press the MENU/ESC key to open the menu. 2. Use S or T to select Data and press the ENTER key. 3.
1. OPERATION 1.18.4 Setting destination waypoint • With the inside-hull transducer, signal attenuation is different between 50 kHz and 200 kHz. Therefore, a fish may not be detected or the fish length indicated may be smaller than actual length. Set a destination waypoint to find range, bearing and time-to-go to that point. Range, bearing and time-to-go (to the waypoint) are shown on the nav data display. • In case of a fish school, echoes overlap one another, so the margin of error will be greater.
1. OPERATION 1.19.3 Displaying fish symbols 1.19.4 Displaying fish info 1. Press the MENU/ESC key to open the menu. 1. Press the MENU/ESC key to open the menu. 2. Use S or T to select Accu Fish and press the ENTER key. 2. Use S or T to select Accu Fish and press the ENTER key. 3. Use S or T to select Fish Symbols and press the ENTER key. 3. Use S or T to select Fish Info and press the ENTER key. 4. Use S or T to select Fish Size or Depth as appropriate and press the ENTER key. 4.
1. OPERATION 1.19.6 1.20 Applying offset to measured fish size Menu Description This section describes menu items not previously mentioned. For the System menu, see chapter 2. If the size of the fish caught differs greatly from the measured value, apply an offset to compensate for the difference. 1.20.1 1. Press the MENU/ESC key to open the menu. Sounder menu 2. Use S or T to select Accu Fish and press the ENTER key. 3. Use S or T to select Fish Size and press the ENTER key. 4.
1. OPERATION 2. Use S or T to select HF Terminal or LF Terminal as appropriate and press the ENTER key. The frequencies preset with Freq Control are shown. Auto shift: Turn the auto shift feature on or off. Select Off to shift the display manually (with S or T). On automatically tracks the bottom echo to keep it on the bottom half of the display. For example, the distance to the bottom is 350 ft and the range is set to 0-60 ft.
1. OPERATION the current display is erased when changing this setting. 1. Choose Freq. Control and press the ENTER key. The contents of the setting window depend on your system configuration. In the example below, a 200 kHz transducer is connected to the HF terminal and a 50 kHz transducer is connected to the LF terminal. 194.0kHz 195.5kHz White Line: The white line helps you distinguish bottom fish from the bottom by tracing a line, in white or color desired, just above the bottom contour.
1. OPERATION 1. Choose Bottom Zone and press the ENTER key. White Marker: Display a specific echo color in white. This feature is useful for emphasizing echoes having a specific strength. 1. Select White Marker and press the ENTER key. Boundary line (top) Bottom display area Boundary line (bottom) 2. Use W or X to select the boundary line to adjust. 3. Use S or T to adjust the boundary line. 2. Use W or X to select high or low frequency as appropriate.
1. OPERATION 1.20.2 Tx/Rx menu resolution, however, detection range is shorter (pulse length is about 1/4 of Std) than the Std setting. Short 2 raises the detection resolution, however, detection range is shorter (pulse length is about 1/2 of Std) than the Std setting. Std is the standard pulse length, and is suitable for general use. Long Increases the detection range but lowers the resolution (about 1/2 compared to the Std pulse length). Manual enables manual setting of pulse length, at "Pulse Length.
1. OPERATION 1.20.3 Display menu Disp Division: Select display division in dual frequency and combination displays (zoom+normal). The choices are shown in the illustration below. : Horizontal split : Vertical split Color Bar: Turn the color bar on or off. Hue: Change the color arrangement. The choices are Std, Hue1 thru Hue6, and Custom. As you move through the selections you can see the color arrangement at the right side of the screen.
1. OPERATION 1.20.4 Stabilizer menu 6. Press T or S to select desired Stabi. Area and then press the ENTER key. The Stabilizer menu compensates for the effects of heaving, and requires a satellite compass. It is inoperative when there is no data from the satellite compass and NMEA 0183 on the NMEA menu is selected to Special. 7. Press the MENU/ESC key twice to close the menu. 1.20.5 Data menu The Data menu sets up data received from external equipment.
1. OPERATION Bearing: Select how to display bearing data, fed from navigator. The choices are true and magnetic. 1. Select Data Box1 and press the ENTER key. Nav Data: Select source of position data, among GPS, Loran C, and Auto. Auto selects navigator in order of accuracy in case of multiple navigators. The order is GPS followed by Loran C. Wind Spd/Dir: Display wind speed and direction data in True or Apparent.
1. OPERATION 1.20.6 NavNet menu By connecting the FCV-1150 to a NavNet 3D unit, you can show the image from the FCV1150 on the display of the NavNet 3D and control some functions from the NavNet 3D. The applicable program version numbers are as follows: MFD8/MFD12: 1950055-02.01 or after MFDBB: 1950064-02.01 or after For operation of the NavNet 3D see its operator’s manual. If you set up for NavNet mode connection at installation, the NavNet menu appears.
2. SYSTEM MENU 2.1 How to Open theSystem Menu 2.2.1 User menu description User Color Press the MENU/ESC key to open the menu, then press S or T to select System. Arrange the display colors to your liking by changing the color arrangement on the color bar. 1. Select User Color and press the ENTER key to show the user color bar. Color currently selected System menu See installation manual. See Chapter 3. Background color 2.2 User Menu 2.
2. SYSTEM MENU User Clutter User 1, User 2 Select the colors to reject with the clutter rejector. Define what to show on the two user display mode screens, selectable with the MODE control. 1. Select User Clutter and press the ENTER key to show the user clutter color bar. Strong color Color 6 Color 5 Color 4 Color 3 Color 2 Color 1 Bkgd color Screen Layout: Select the screen layout, from among the following eight choices.
2. SYSTEM MENU Pulse, Short 1; Echo Stretch, ON, and Smoothing, Off. Deep Sea is the same as Normal. About the mix display The mix display compares echo intensity between low and high frequencies, and displays echoes from tiny fish in discriminative colors. This is done by utilizing the fact that tiny fish return a stronger echo against a high frequency rather than a low frequency. It works as follows. Nav Data Disp: Turn the nav data display On or Off and select character size, from large or small.
2. SYSTEM MENU 3. Use W or X to select the nav data item to display. Availability depends on how much nav data is displayed, as shown below.
2. SYSTEM MENU 2.3 Range Menu Zoom Range Zoom range sets the range for the marker zoom and bottom zoom displays. Press S or T to set zoom range as desired. For the horizontal split screen, the range is one-half the value set. The Range menu is where you can pre-set basic ranges, zoom range, bottom lock range, and turn independent range adjustment on or off. The default values are suitable for most fishing applications, however you may wish to change pre-set ranges to suit your needs.
2. SYSTEM MENU 2.4 Key Menu 2.6 The Key menu selects the function for the FUNCTION control and turns the key beep on or off. Units Menu The Units menu lets you select the unit of measurement for depth, temperature, speed, wind, and distance, from the choices shown below. * 2.5 Language Menu * Not adjustable when NavNet mode is active. The Language menu selects the language to use. Select Language and press the ENTER key. Use S or T to select language.
2. SYSTEM MENU Sound Speed Bottom Level Adjust the sound velocity of the Tx/Rx signal if the depth indication is incorrect, because of water temperature or salinity density. In the default bottom level setting (0), the equipment judges consecutive strong echoes to be bottom echoes. If, in that setting, the depth indication is unstable, adjust the bottom level. If vertical lines extend upward from the bottom echo in the bottom lock display, lower the bottom level to erase the vertical lines.
2. SYSTEM MENU [HF], [LF] Draft: The default depth display shows the distance from the transducer. If you would rather show the distance from the sea surface, set your ship's draft. Gain ADJ: If the gain is too high or too low, or there is a difference in the gain between the low and high frequencies, you can compensate for it here.
3. MAINTENANCE & TROUBLESHOOTING 3.2 WARNING ELECTRICAL SHOCK HAZARD Do not open the equipment. Care of Display Unit Dust or dirt on the cabinet can be removed with a soft, dry cloth. For stubborn dirt, waterdiluted mild detergent can be used. In this case, wipe the cabinet with a dry cloth after using detergent. Do not use solvents such as thinner, acetone or benzene to clean the unit. They can remove paint and markings. Hazardous voltage exists inside the equipment.
3. MAINTENANCE & TROUBLESHOOTING 3.5 Battery Voltage Alert A battery icon appears when the battery voltage is too high or too low. Check the battery when a battery icon appears. Icon Remedy Extreme interference or noise • Check if transducer is too close to engine. • Check if unit is properly grounded. • Check if other echo sounders of same frequency as own are being operated nearby. Meaning Voltage is lower than 10 VDC. If the voltage goes below 9 V, the equipment automatically shuts itself off.
3. MAINTENANCE & TROUBLESHOOTING 3. Press the ENTER key. ROM check SDRAM check Water temperature* Speed* 5 SDRAM NMEA check Voltage* 89321 7587 Color bar For factory use 0252359-XX.xx 0252360-XX.xx 0252359-XX.xx 0252360-XX.xx XX Control test XX=Program no. xx=Program version no. * Updated every second. 4. Press S to select Test and then press the ENTER key. The self test results appear together with the key test screen, as shown above.
3. MAINTENANCE & TROUBLESHOOTING 3.8 LCD Test 3.9 The LCD test checks for proper display of all colors. To stop the test at any time, press the MENU/ESC key. You may wish to restore default settings to start afresh. 1. Press the MENU/ESC key to open the menu. 1. Press the MENU/ESC key to open the menu. 2. Press T to select Tests to show the Tests menu and press the ENTER key. 2. Press T to select Reset at the bottom of the menu bar to show the Reset menu. 3.
APPENDIX 1 MENU TREE Sounder MENU key TxRx Pic. Advance (4/1, 2/1, 1/1, 1/2, 1/4, 1/8, 1/16, Stop) Zoom Mode (Bottom Lock, Bottom Zoom, Marker Zoom, Discrim1/2*, Discrim1/3*) Free Shift (Off, On) * Not available when Auto Shift (Off, On) NavNet menu is set for NavNet. Interference (Off, Low, Medium, High, Auto) Freq Choice (Settings according to transducer connected.) Freq Control (Settings according to transducer connected.
APPENDIX 1 MENU TREE 1 Display A-Scope (Off, Normal, Peak) Depth Size (Off, Small, Middle, Large) Depth Scale (Right, Center, Off) Zoom Marker (Off, On) Temp Graph (Off, Narrow, Wide) Temp Graph Color (Std, White, Red, Black, Yellow) Pic. Adv.
APPENDIX 1 MENU TREE 2 Data Go to WPT (Off, Destination 1 - 20) WPT List Data Box 1 (Speed (SOG)(Off, On), Speed (STW)(Off, On), Depth (Off, On), Range(Off, On), Bearing (Off, On), Position (Off, On), Wind (Off, On), Heading (Off, On), Course (Off, On), Barom Press (Off, On), Temperature (Off, On), Time to Go (Off, On), Trip Meter (Off, On), Odometer (Off, On), XTE (Off, On), TD (Off, On), Timer (Off, On), Scroll Time (Off, On), Battery, Switching Cycle (2s, 4s, 6s, 8s, 10s, 15s, 20s, 30s) Data Box 2 (Sam
APPENDIX 1 MENU TREE 3 Key Lang Units NMEA FUNC Key (Pic. Advance, Interference, Color Erase, Clutter, White Line, White Marker, TVG, STC, Bottom Zone, A-Scope, Split Range, Auto Shift, Freq Choice, Freq Control, Gain/Mode*1 ) Key Beep (Off, On) Language (English, OTHERS) Depth (m, ft, fa, HR, pb)*2 Temp (°C, °F) Speed (kt, km/h, mph) Wind (kt, km/h, mph, m/s) Distance (nm, km, sm) NMEA0183 (Ver 1.5, Ver 2.0, Ver 3.
APPENDIX 2 SCREEN LAYOUT The screen may be divided as desired with Screen Layout on the User menu.
APPENDIX 2 SCREEN LAYOUT Two-way split H Z O O M H HIGH NORMAL N O R M A L L Z O O M HIGH ZOOM L [HF]: Zm/Nor L H N O R M A L N O R M A L N O R M A L HIGH NORMAL LOW NORMAL L H Z O O M Z O O M MIX NORMAL N O R HIGH NORMAL L N O R M A L 2 N O R M A L 1 LOW ZOOM N O R M A L M I X MIX NORMAL N O R [HF]+[MIX] H HIGH ZOOM [LF]:Zm+[HF]:Zm M I X H LOW ZOOM [LF]: Zm/Nor [LF]+[HF] H LOW NORMAL N O R M A L LOW NORMAL [LF]+[MIX] H NORMAL1 H NORMAL2 H2+H1 L L N O R M A L 2
APPENDIX 2 SCREEN LAYOUT Three-way split L N O R M A L H HIGH NORMAL H N O R M A L Z O O M HIGH ZOOM L L H Z O O M N O R M A L N O R M A L LOW NORMAL [LF]+[HF]: Zm/Nor L H N O R M A L N O R M A L M I X N O R HIGH NORMAL LOW NORMAL LOW ZOOM [LF]: Zm/Nor+[HF] MIX NORMAL HIGH NORMAL LOW NORMAL [LF]+[HF]+[MIX] Four-way split LOW NOR LOW ZOOM HIGH NOR HIGH ZOOM [LF]: Zm/Nor+[HF]: Zm/Nor AP-7
APPENDIX 3 SCREEN DIVISION The display may be divided vertically and horizontally as shown below, with DIsplay in the DIsplay menu.
APPENDIX 4 NAVNET MODE AND FUNCTION AVAILABILITY Function (Name at NavNet in parentheses if different) Gain Mode*1 Auto Range Auto Gain (Auto Mode) Auto Gain Offset (Auto Gain Calib) Shift Shift*2 Auto Picture Advance Zoom Mode*3 Interference Rejection Frequency Control Color Erase Clutter White Marker TVG STC Smoothing*4 (High Resolution) Bottom Zone (Bottom Range Shift) TX Power (Transmission Power) TX Rate (Transmit Rate) TX Pulse Pulse Length RX Band Available with unit selected at Gain/Mode Availabl
APPENDIX 4 NAVNET MODE AND FUNCTION AVAILABILITY Function (Name at NavNet in parentheses if different) A-scope Zoom Marker (Zoom Reference Lines) Temp Graph Background (Background Color Colors (Echo Color Levels) ACCU-FISH Fish Info (Info Size) Fish Symbols Fish Size Stabilization (Heaving) Fish Alarm/ Fish B/L Alarm relationship Fish Alarm Level Zoom Range (Zoom Range Span) Split Range Depth Unit (Depth) Sound Speed Bottom Level Draft Gain Adjust Zero Line, Zero Line Area Available with unit selected at
FURUNO FCV-1150 SPECIFICATIONS OF COLOR LCD SOUNDER FCV-1150 1 1.1 1.2 1.3 1.4 1.5 ECHO SOUNDER TX frequency Output power Power reduction TX rate Pulselength 2 2.1 2.2 2.3 2.4 2.5 2.6 2.7 DISPLAY UNIT Display mode Brilliance Echo color Background color Range shift Display mode Zoom display 2.8 Picture advance speed 2.9 Alarm 2.10 Noise limiter 2.11 Automatic indication 2.12 Additional function 28/38/50/68/82/88/107/150/200 kHz, select 2 channels 1, 2 or 3 kW Auto/ 10 to 100%, 10% step Max.
INDEX A Accu Fish enabling/disabling .....................................18 fish info ......................................................19 fish size offset ...........................................20 fish symbols ..............................................19 Alarms activating ...................................................13 arrival ........................................................13 bottom .......................................................13 fish (bottom lock) ....................
INDEX Smoothing ...................................................23 Sound speed offset .....................................35 Sounder menu ............................................20 Speed alarm................................................13 Speed offset ................................................35 Speed source ..............................................27 Split range ...................................................33 Stabilizer menu ...........................................