Technical data

Instructions for Use
6 - English A26361-K1001-Z290-2-8N19, edition 1
Table 3:Guidelines and manufacturer's declaration – Electromagnetic
fault tolerance
The systems discussed in this manual are intended for use in the type of electromagnetic
environment specified below. The customer or user of the device or system should insure that it is
used in such an environment.
Interference
immunity tests
IEC 60601
test level
Conforman
ce level
Electromagnetic environment - guidelines
Portable and mobile wireless equipment should
not be used at a distance to the device or to the
system that is less than the recommended safe
distance. This is calculated according to the
equation dependent on the broadcast frequency.
Recommended safe distance:
Conducted RF
interference as
per
IEC 61000-4-6
3 V
rms
150 kHz to
80 MHz
3 V
P
V
3.5
d
1
Radiated RF
interference as
per
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
3 V/m
P
E
3.5
d
1
for 80 MHz to 800 MHz
P
E
7
d
1
for 800 MHz to 2.5 GHz
with P as the maximum nominal power output of
the sender in watt (W) according to specifications
of the manufacturer of the sender and d as the
recommended safe distance in meters (m).
The field strength of stationary radio transmitters
should be less than the conformance level
b
for all
frequencies in accordance with on-site
a
testing.
In the vicinity of devices that bear following
symbol interferences are possible.