Common Correlation Detection Method
① Turn on the pre-amplifiers.
② Turn on the main unit.
③ Verify that the pre-amplifier icons are lit.
④ Check the battery level and date and time of
the main unit and pre-amplifiers.
To change the date and time, see “3-6 Date
and Time Adjustment” on p. 25.
⑤ Press the pre-amplifier switch
.
⑥ Set the pre-amplifier operation mode, filter,
sensitivity, GPS, and other conditions.
For details on configuring the pre-amplifier,
see "Pre-Amplifier Settings" under "4-7
Explanation of Equipment Settings and
Operations" on p. 35.
⑦ Install the pre-amplifier on the valve etc. at
the installation location.
⑧ Press and hold the sensitivity up or down
switch on the pre-amplifier to perform
automatic adjustment.
If you are unable to press the sensitivity
switch on the pre-amplifier, turn on
"Auto-sens." in the pre-amplifier settings on
the main unit and then tap Send settings.
⑨ Verify that the pre-amplifier icon on the
main unit is lit, and then press the filter
switch
.
⑩ Configure the Td range, automatic filter, and
other settings.
For details on configuring the filter, see
"Filter Setting" under "4-7 Explanation of
Equipment Settings and Operations".