Hearing Aid User Manual

Automated Test Sequences 119
Figure 5.2.3—ANSI 92 test results
5.2.4 Viewing ANSI measurements in the Coupler screen
It’s possible to copy ANSI response curves to the Coupler screen where they
can be compared to other response curves. To do this:
1. Run the ANSI 92 test sequence.
2. Press [F6].
3. Use [
] to select Store Crvs Coup, if necessary.
4. Press [>] to store the curves and close the menu. The following is a list of
where the curves are stored in the Coupler screen. Any previous measure
-
ments stored in those curve locations will be overwritten.
50 dB SPL—CRV3
60 dB SPL—CRV4
70 dB SPL—CRV5
80 dB SPL—CRV6
90 dB SPL—CRV7
5. Press [EXIT] and [F1] to enter the Coupler screen and view the ANSI
response curves.