Table 6. Troubleshooting Guidelines
CONDITION POSSIBLE CAUSE POSSIBLE CORRECTIVE ACTION
Power LED on Low SAFECONNECT Replace Belt Bridge batteries
SAFECONNECT Belt batteries
Bridge is RED
POWER LED on Sirius Wireless Link Replace Wireless Link battery
SAFECONNECT Belt battery is low
Bridge is ORANGE
POWER LED on Failure to press the Press the ON/OFF button until all
SAFECONNECT ON/OFF button for three LEDs turn green
Belt Bridge is OFF a full second
Low batteries Replace batteries
Unit inoperative Return unit to factory for repair
COMM LED on Absence of Bluetooth Momentarily press the LINK button
SAFECONNECT Belt link between Wireless while both devices are within range
Bridge is ORANGE Link and Belt Bridge
Reset Wireless Link by pressing
LINK button for five seconds;
then, attempt to establish link
Replace Wireless Link battery and
attempt to establish LINK
Absence of link between Wireless Link should be seated or
Wireless LINK and mated properly with Sirius
Sirius instrument instrument in boot
At cold temperatures Cycle power on Sirius instrument to
(<0°C) loss of communi- reinitialize communication.
cation can occur Wear Sirius instrument close to body
to keep it warm.
Sirius instrument is OFF Turn instrument ON, wait until
start-up routine is complete, and
attempt to establish LINK
Direct, bright sunlight on Ensure that provided MSA label is
IRDA communication installed at top of Sirius unit and
from Sirius to Sirius providing sunlight shield
Wireless Link can cause
loss of communication
COMM LED on SAFECOM unit is OFF Turn SAFECOM unit ON
SAFECONNECT
Belt Bridge is RED
(Belt Bridge is not
connected to
SAFESITE network)
Incorrect network ID Return unit to factory for proper
and/or frequency key on configuration
Belt Bridge
Belt Bridge out of Belt Bridge will automatically
network range reconnect when it is in network
range
Radio inoperative Return unit to factory for repair