Caution! Only process and heat up any media that has
a flash point higher than the adjusted safe temperature
limit that has been set.
The safe temperature limit must always be set to at least
50°C lower than the fire point of the media used.
Beware of hazards due to:
- flammable materials
- combustible media with a low boiling temperature
- glass breakage
- incorrect container use
- overfilling of media
- unsafe condition of container
The machine may heat up when in use. Don’t use the
machine in explosive atmospheres with hazardous
substances.
Process pathogenic materials only in closed containers
under a suitable extractor hood.
Only process media that will not react dangerously to
the extra energy produced in other ways, e.g. through
light irradiation.
The external temperature sensor PT1000 must always
be inserted in the media when connected and ensure
it’s inserted in the media to a depth of at least 20mm.
Accessories must be securely attached to the machine
and can’t come off by themselves.
Always disconnect the plug before fitting accessories.
The machine can only be disconnected from the mains
supply by pulling out the mains plug or the connector
plug.
When using PTFE-coated magnetic bars, the following