3
• Caution! Only process and heat up any media that has
a flash point higher than the adjusted safe temperature
limit that has been set.
• The safe temperature limit must always be set to at least
50°C lower than the fire point of the media used.
• Beware of hazards due to:
- flammable materials
- combustible media with a low boiling temperature
- glass breakage
- incorrect container use
- overfilling of media
- unsafe condition of container
• The machine may heat up when in use. Don’t use the
machine in explosive atmospheres with hazardous
substances.
• Process pathogenic materials only in closed containers
under a suitable extractor hood.
• Only process media that will not react dangerously to
the extra energy produced in other ways, e.g. through
light irradiation.
• The external temperature sensor PT1000 must always
be inserted in the media when connected and ensure
it’s inserted in the media to a depth of at least 20mm.
• Accessories must be securely attached to the machine
and can’t come off by themselves.
• Always disconnect the plug before fitting accessories.
• The machine can only be disconnected from the mains
supply by pulling out the mains plug or the connector
plug.
• When using PTFE-coated magnetic bars, the following