User Manual

10
- Check if the fuse is broken or loose
Fault in power on self test
- Power off the device and then restart
Actual speed can’t reach the set value
- Media in high viscosity may probably cause abnormal
speed reduction of the motor
The device can’t be powered off when switched off
- Check if the residual heat warning function is still on
and the hotplate temperature is above 50°C (LED still
works and “Hot” flashes)
Error codes
Error code
Error description Solutions
Solutions
Er1
Short circuit in the external
temperature sensor PT1000
Replace PT1000 and
re-start the device
Er2
Open circuit in the internal
temperature sensor PT1000
Re-start the device
Er3
Short circuit in the internal
temperature sensor PT1000
Re-start the device
Er4
Excessive temperature of the
device
Power off, and then
re-startthe device after
cooling down
Er5
Motor failure
Reduce the sample
If the instrument still can’t work properly, please contact
manufacturer/supplier for technical support.