User Manual
第 2 页 共 4 页
四、 Detailed operation instructions
1. Temperature calibration
Note: The initial use of the machine must be calibrated once,
otherwise problems such as the inability to save parameters under
normal use may occur.
a) After power on, long press the encoder button > 3S, the system
enters the calibration mode. At this time, the display only shows
the temperature value and the clock value disappears, ensuring
that the calibration mode is entered.
b) Press PRO key to select setting calibration point 40/60/80/100 (if
the temperature value is not the specified value, adjust the
encoder to the specified value), or you can directly adjust the
encoder to the specified calibration point value without pressing
the PRO key.
c) Add the high-boiling solution (greater than 100°, such as
glycerin) to the upper BLOCK hole and immerse the standard
thermometer probe in the BLOCK solution. Make sure that the probe
is located in the hole in or near the geometric symmetry of the
BLOCK.
d) Press the RUN/STOP key to start warming up and the system status
Encode
r
Button
Cover
BLOCK
Display
yy