Service Manual
{SYSTEM}
System contains the following parameters, separated by tab (\t) or newline (\n)
id: 1835060 A unique number identifying the master processor
digital filter: 1 System settings for digital filter (50Hz Cut-off). 0 is off, 1 is on
digital AC: 1 System settings for digital AC filter. 0 is off, 1 is on
mute vectors: 0 0 sends all vector data, 1 mutes the vector data
menu active: 1 1 activates the menu and allows you to send short commands for start, stop
and help menu. 0 deactivates the menu
50hz filter: 1 50Hz analogue filter. 0 deactivates and 1 activates
220nf cap: 0 Filter for active pipe-tracker, e.g. TSS440. 0 deactivates, 1 activates
100x gain: 1 Gain setting on FiGS 0 is 10x gain and 1 is 100x gain
rotor state: 1 1 is measure, 2 is impedance
motor state: 0 Motor state, 0 is off, 1 is on
no rotation mode: 0 0 no-rotation mode deactivated, 1 no-rotation mode active
electrode distance: 7.00 Distance between electrodes in [cm]
use dir cal: 1 Direction calibration: 0 deactivated, 1 active
dir cal: -72.13 active direction calibration angle
use level calibration: 1 Level calibration, 0 is deactivated, 1 is active
level calibration: 147.79 Level calibration angle
use mounted angle: 1 use mounted angle, 0 is deactivated, 1 is active
mounted angle: 147.71 mounted angle
target RPM: 500 system target RPM
use relative cal: 1 use relative cal, 0 is deactivated, 1 is active
relative cal deg: 41.135 relative cal. angle
relative cal amp: 1.053 relative cal. amplitude
use absolute cal: 1 use absolute calibration, 0 is deactivated, 1 is active
absolute cal factor: 1.010 absolute calibration factor
use geometry factor: 1 use geometry factor, 0 is deactivated, 1 is active
geometry factor: 0.714 geometry factor value