Owner's Manual
Table Of Contents
EN-5
If you want to turn o an alarm, nd the alarm number by
pressing SET in the setting mode, press M to change from “ON’
to “OFF”.
At the time of your alarm, the meter will beep and automatically
turn on. You can press M to silence the alarm and insert a test
strip to begin testing. If you do not press M, the meter will beep
for 2 minutes then switch o. If you do not want to test at this
time, press M to switch o the meter.
Press SET to turn o the meter. The meter repeats all your
settings before shutdown.
Congratulations! You have completed all settings!
Note:
• These parameters can ONLY be changed in the setting mode.
• If the meter is idle for 3 minutes during the setting mode, it will
switch o automatically.
Before Testing
Control Solution Testing
FORA Control Solution contains a known amount of glucose that
reacts with test strips and is used to ensure your meter and test
strips are working together correctly.
Do a control solution test when:
9 you suspect the device or test strips are not working properly.
9 your blood glucose test results are not consistent with how
you feel, or if you think the results are not accurate,or
9 you have dropped or think you may have damaged the meter.
Performing a Control Solution Test
Step 1: Insert the test strip to turn on the meter.
Insert the test strip into the meter. Wait for the meter to display
the test strip and blood drop symbols.