Owner's Manual
Line
Input Response
•
Gain
range=
-20dB to +10dB in
10dB
steps+
20
dB of
variable
gain
Input
Impedance=
10k0
from
10Hz to 200kHz
Noise at main
output
with
gain at
unity
[OdB)
= -96dBu
measured
with
a 22Hz/22kHz band-pass
filter
Signal to noise
ratio
relative
to
max
headroom
[24dBu)=120dB
•
Signal to noise
ratio
relative
to OdBFS [+22dBu) = 118dB
•
THO
at
unity
gain
[OdB)
= 0.001%
measured
with
a
OdBu
input
signal
and
with
a 22Hz/22kHz
band-pass
filter
•
Frequency Response at
unity
gain
[OdB)
= -0.3dB down at 10Hz and
-3dB
down at 200kHz
Instrument Input Response
Gain
range=
10dB to 40dB
continuously
variable
Input Impedance:
High=
1MO
Low=
300k0
•
Noise at
minimum
gain
[+10dB)
= -92dBu
measured
with
a 22Hz/22kHz band-pass
filter
•
Noise at
maximum
gain [+40dB) = -62dBu
measured
with
a 22Hz/22kHz band-pass
filter
•
THO
at
minimum
gain [+10dB) = 0.001% measured
with
a 10dBu
input
signal
and
with
a 22Hz/22kHz
band-pass
filter
•
Frequency Response at 10dB gain
with
-10dB
input=
10Hz-100kHz+/-
0.6dB
•
Frequency Response at 40dB gain
with
-40dB
input=
-2.5dB down at 10Hz and
OdB
at 100kHz
High Pass Filter
•
Roll
off=
18dB per octave 3 pole
filter
•
Fixed Frequency 75Hz
measured
at the 3dB down point
Meters
•
Moving
coil[MC)
meter
is
factory
calibrated
to
OVU
= +4dBu
with
1kHz sinewave. With the
VU
Cal
button
pressed, the
meter
can
be
adjusted
on
the
rear
panel to
allow
OVU
to
equal+
10dBu to +26dBu
with
the
centre
detent being
equal
to +22dBu
• Peak
LED
meters
calibrated
in the
detent
position
for
OdBFS = +22dBu,
calibration
is
adjustable
on the
rear
panel to
allow
OdBFS
to equal +10dBu to +26d8u
Routing
for
MC
and Peak1
meter
is
after
the
HPF,
pre
insert
send
or
switched
post
insert
return.
Peak2
is always pre
ADC
channel2,
which
can
be
fed by
external
input
or
Instrument
input.
LED
levels
are as
follows
when
peak
calibration
is set to
center
detent
on
the
rear
panel. [This is
when
using the
internal
ADC).
0
+22dBu
-2
+20dBu
-6
+16dBu
-12
+10dBu
-18
+4dBu
-42
-20dBu
20