Instruction Manual

Beta FLUO BASIC Manual 2.02
3/29/2022
Press the RESET button and keep the right key depressed during reset
to perform a FACTORY RESET. The factory Reset will reset all settings and
configuration to its defaults.
Configure the device to your needs
The FLUO can be configured to your needs in a special Configuration mode.
Press and hold the right key for at least 5 seconds to enter Configuration
mode
The configuration mode is indicated by a ‘~’ in the left top corner of the LCD.
Two settings can be configured:
a) Average Solid Reference
b) Color Mode or Intensity Mode
Click the right key to adjust the Average Solid Reference settings and
Color or Intensity mode selection.
a) Use an average solid reference
The FLUO can be configured to calculate the average of a series of up to 10
readings on your reference patch. The number of readings that need to be
measured to calculate the reference can be set in the Configuration mode.
Averaging multiple readings can be useful when there is unevenness or
variability in the reference patch print quality.
Click the up key on the device to increment the number of readings
Click the down key to decrement the number of readings