Data Sheet
3 Fluke Corporation Ti400, Ti300 and Ti200 Infrared Cameras with LaserSharp
®
Auto Focus
Detailed specifications
Ti400 Ti300 Ti200
Battery
Batteries (field-replaceable, rechargeable) Two lithium ion smart battery packs with five-segment LED display to show charge level
Battery life Four+ hours continuous use per battery pack
Battery charge time 2.5 hours to full charge
Battery charging system Two-bay battery charger or in-imager charging. Optional 12 V automotive charging adapter
AC operation AC operation with included power supply (100 V AC to 240 V AC, 50/60 Hz)
Power saving User selectable sleep and power off modes
Temperature measurement
Temperature measurement range (not calibrated below -10 °C) -20 °C to +1200 °C
(-4 °F to +2192 °F)
-20 °C to +650 °C
(-4 °F to +1202 °F)
Accuracy ± 2 °C or 2 % (at 25 °C nominal, whichever is greater)
On-screen emissivity correction Yes (both value and table)
On-screen reflected background temperature compensation Yes
On-screen transmission correction Yes
Color palettes
Standard palettes 8: Ironbow, Blue-Red, High Contrast, Amber, Amber Inverted, Hot Metal, Grayscale, Grayscale Inverted
Ultra Contrast
TM
palettes 8: Ironbow Ultra, Blue-Red Ultra, High Contrast Ultra, Amber Ultra, Amber Inverted Ultra,
Hot Metal Ultra, Grayscale Ultra, Grayscale Inverted Ultra
General specifications
Color alarms (temperature alarms) High-temperature and low-temperature
Infrared spectral band 7.5 μm to 14 μm (long wave)
Operating temperature -10 °C to +50 °C (14 °F to 122 °F)
Storage temperature -20 °C to +50 °C (-4 °F to 122 °F) without batteries
Relative humidity 10 % to 95 % n on - c on den s i n g
Center-point temperature measurement Yes
Spot temperature Hot and cold spot markers
User-definable spot markers 3 user-definable spot markers
Center box Expandable-contractible measurement box with MIN-MAX-AVG temp
Safety IEC 61010-1: Overvoltage category II, Pollution Degree 2
Electromagnetic compatibility IEC 61326-1: Basic EM environment. CISPR 11: Group 1, Class A
Australian RCM IEC 61326-1
US FCC CFR 47, Part 15 Subpart B
Vibration 0.03 g2/Hz (3.8 g), 2.5 g IEC 68-2-6
Shock 25 g, IEC 68-2-29
Drop Engineered to withstand 2 meter (6.5 feet) drop with standard lens
Size (H x W x L) 27.7 cm x 12.2 cm x 16.7 cm (10.9 in x 4.8 in x 6.5 in)
Weight (battery included) 1.04 kg (2.3 lb)
Enclosure rating IEC 60529: IP54 (protected against dust, limited ingress; protection against water spray from all directions)
Warranty Two-years (standard), extended warranties are available
Recommended calibration cycle Two-years (assumes normal operation and normal aging)
Supported languages Czech, Dutch, English, Finnish, French, German, Hungarian, Italian, Japanese, Korean, Polish, Portuguese,
Russian, Simplified Chinese, Spanish, Swedish, Traditional Chinese, and Turkish