Cut Sheet

2 Fluke Corporation Fluke T+ and T+PRO Electrical Testers
Specifications
Temperature
Operating -10 °C to 55 °C (14 °F to 131 °F)
Storage
-30 °C to 60 °C (-22 °F to 140 °F)
Relative humidity 0 °C to 30 °C (32 °F to 86 °F): 90 %
30 °C to 40 °C (86 °F to 104 °F): 75 %
40 °C to 50 °C (104 °F to 122 °F): 45 %
Altitude
Operating 2,000 m (approx. 6,561.6 ft)
Storage 10,000 m (approx. 32,808.4 ft)
Battery type/life AAA (2); 40 hours
Shock, vibration Sinusoidal vibration per MIL-PRF-28800F for a Class 2 instrument
Safety ISA-82.02.01 (IEC 61010-1 Mod) Second Edition, CAN/CSA-C22.2 No. 61010-1
Second Edition, IEC 61010 - 1 to 1000 V CAT III/600 V CAT IV
Compliance to NFPA 70E
1
AC Bandwidth 45 Hz to 66 Hz
Maximum working voltage 1000 V ac/dc
Maximum measurable voltage 600 V ac/dc
Duty cycle Indefinitely for voltages up to 240 V.
For voltages between 240 V and 600 V the duty cycle is 30 s on / 300 s off.
For voltages above 240 V the tester must connect to a voltage source only for a
MAXIMUM of 30 s and then disconnect for a MINIMUM of 300 s.
Features and Benefits contd.
Replaceable test probes for added flexibility and
convenience
Convenient probe storage on the back
Probe holder for easier testing
Hazardous voltage warning light for added user
protection
Test lead strap included for better test lead
management
AC/DC Voltage: 12 V – 600 V on LED bar, 10.2 V
– 600 V on LCD (LCD on T+PRO only)
Auto-off saves battery power
2-year warranty
Added functionality available only
on the T+PRO
Backlit LCD display for easier viewing
Display resolution of 0.1 V, ideal for
troubleshooting low-voltage lighting
Resistance (Ohms) to 9.99 kOhms
Rotary Field Indication
Display Hold for easier viewing in hard-to-test
areas
1
When used properly per Article 110.9 Use of Equipment (A) Test Instruments and Equipment.