Datasheet

ScopeMeter 190 Series II Fluke Corporation 7
FlukeView
®
ScopeMeter
®
Software
System requirements
Microsoft
®
Windows
®
XP and beyond
CD-ROM drive
One free USB port
Supported Instruments
With the new release V5, the following typenumbers
are supported:
Fluke 190C-series (225C, 215C, 199C, 196C, 192C,
using an OC4USB or PM9080 interface cable);
Fluke 190B-series (199B, 196B, 192B, using an
OC4USB or PM9080 interface cable);
190-series II (190-204 and 190-104, using
USB-cable);
120-series (123, 124, 125, using an OC4USB or
PM9080 interface cable).
FlukeView ScopeMeter software helps you
get more out of your ScopeMeter:
Store instrument’s screen copies on the PC, in color
Copy screen images into your reports and
documentation
Capture and store waveform data from your
ScopeMeter on your PC
Create and archive waveform references for
automatic or visual comparison
Includes waveform analysis, e.g. FFT spectrum
analysis
Copy waveform data into your spreadsheet for
detailed analysis
Use cursors for parameter measurement
190C Series 190 Series II
199C, 196C, 192C, 215C, 225C, 190-204, 190-104
Battery operating time (with backlight low) > 3½ hours Up to 7 hours using BP291 (included)
Battery charging time 4 hours 5 hours
Battery power saving functions Auto ‘power down’ with adjustable
power down time.
On-screen battery power indicator.
Auto ‘power down’ with adjustable
power down time; Auto ‘Display off’ with
adjustable power down time; On-screen
battery power indicator.
Safety
Compliance EN61010-1-2001, Pollution Degree 2;
UL61010B, with approval;
CAN/CSA C22.2, No. 61010-1-04, with
approval; ANSI/ISA-82.02.01
EN61010-1-2001, Pollution Degree 2;
CAN/CSA C22.2, No. 61010-1-04, with
approval; UL61010B; ANSI/ISA-82.02.01
Environmental
Operating temperature 0 °C ~ +50 °C 0 °C ~ +40 °C incl. battery
+40 °C ~ +50 °C excl. battery
Storage temperature -20 °C ~ +60 °C
Humidity +10 °C ~ +30 °C: 95 % RH non-condensing
+30 °C ~ +40 °C: 75% RH non-condensing
+40 °C ~ +50 °C: 45% RH non-condensing
Maximum operating altitude 3,000 m (10,000 feet) Up to 2,000 m (6666 ft) for CAT IV 600 V,
CAT III 1000 V; up to 3,000 m (10,000 ft) for
CAT III 600 V, CAT II 1000 V
Maximum storage altitude 12 km (40,000 ft)
Electro-Magnetic-Compatibility (EMC) EN 61326-1 for emission and immunity EN 61326-1 (2005-12) for emission and
immunity
Interface
Optical port in instrument transfers
instrument settings, screen images and
waveform data, compatible with FlukeView®
software for Windows®, via optional OC4USB
or PM9080 (optical to electrical interface
cable)
Two USB ports provided. Ports are fully
insulated from instruments fl oating
measurement circuitry. USB-host port directly
connects to external ash memory drive
for storage of waveform data, measurement
results, instrument settings and screen
copies. A mini-USB-B is provided which
allows for interconnection to PC for remote
control and data transfer under PC-control.
Warranty
Three-years (parts and labor) on main instrument, one-year on accessories
Probe calibration output (through DMM-input banana connectors) Dedicated probe-cal output with reference
contact provided, fully insulated from any
measurement input channel