Datasheet
4 Fluke Corporation ScopeMeter 190 Series II
Bus Health Test Mode (225C and 215C models only)
Bus Health automatically analyzes the electrical signals on the industrial bus system to measure individual parameters and to give waveform
information. Automatically compares the measurement results to preset values and present ‘good,‘weak’ or ’false’ indicator with each parameter.
Bus types and reference standards used
•
AS-i (EN50295, 166 kb/s);
•
CAN-bus (ISO-11898, up to 1 Mb/s);
•
Modbus (EIA-232 up to 115 kb/s and EIA-485 up to 10 Mb/s);
•
Foundation Fieldbus H1 (61158 type 1, 31.25 kb/s) ;
•
Profibus DP (EIA-485 up to 10 Mb/s) and PA (61158 type 1 31.25 kb/s);
•
Ethernet [10Base2 (coaxial) and 10BaseT (UTP)], 10 Mb/s;
•
Ethernet 100BaseT (100 Mb/s);
•
RS-232 (EIA-232, up to 115 kb/s);
•
RS-485 (EIA-485, up to 10 Mb/s).
Measured parameters (where applicable) Bias voltage level, signal amplitude, pulse width or baud rate, risetime, fall time, jitter, signal
distortion, noise HF, noise LF, in-band noise
Meter Mode
190C Series 190 Series II
199C, 196C, 192C, 215C, 225C, 190-204, 190-104
Meter inputs
Via 4 mm banana inputs, fully isolated from
scope inputs and scope ground
Up to four automatic meter measurements
can be made at the same time, using the
oscilloscope input channels
The specifi ed accuracy is valid over the temperature range 18 °C to 28 °C (65 °F to 82 °F).
Add 10 % of specifi ed accuracy for each degree C below 18 °C or above 28 °C.
Maximum resolution
5,000 counts 999 counts
Meter input impedance
1 MΩ ± 1 % // 10 pF ± 2 pF (thru scope channel:)
1 MΩ ± 1 % // 14 pF ± 2 pF
Advanced meter functions
Auto/manual ranging, relative measurements (Zero reference), TrendPlot recording
Vdc, Vac, Vac+dc
Vdc accuracy ± (0.5 % + 5 counts) ± (1.5 % + 5 counts)
Vac true rms accuracy
15 Hz to 60 Hz: ± (1 % + 10 counts) ± (1.5 % + 10 counts)
60 Hz to 1 kHz: ± (2.5 % + 15 counts)
60 Hz to 20 kHz: — ± (2.5 % + 15 counts)
Vac+dc true rms accuracy
15 Hz to 60 Hz: ± (1 % + 10 counts) ± (1.5 % + 10 counts)
60 Hz to 1 kHz: ± (2.5 % + 15 counts)
60 Hz to 20 kHz: — ± (2.5 % + 15 counts)
Voltmeter ranges
500 mV, 5 V, 50 V, 500 V, 1,000 V
Ohms
Ranges 500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ —
Accuracy ± (0.6 % + 5 counts) —
Other meter functions
Continuity Beeper on < 50 Ω (± 30 Ω) —
Diode test Up to 2.8 V —
Amps Adc, Aac, Aac + dc using an optional current clamp or shunt.
Scaling factors: 0.1 mV/A, 1 mV/A, to 100 V/A and 400 mV/A
Temperature With optional accessories. Scale factors 1 °C/mV or 1 °F/mV